Development and Clinical Implementation of Next Generation Very Light Weight and Extremely Flexible Receiver Arrays for Pediatric MRI

Shreyas Vasanawala, Robert Stormont, Scott A. Lindsay, Thomas Grafendorfer, Joseph Y. Cheng, John M. Pauly, Greig Scott, Jorge X. Guzman, Victor Taracila, Dan Chirayath, Fraser RobbarXiv 2017 · DOI

Abstract

Traditional MRI receiver coils are often heavy, rigid, and designed for adult anatomy, posing significant challenges for pediatric imaging. This study introduces a next-generation, 16-element highly flexible pediatric coil array (AIR technology) weighing only 480 grams. The array utilizes novel conductors and miniaturized electronics to maintain element decoupling even under extreme flexion. In a prospective clinical trial of 24 pediatric subjects, the array demonstrated high diagnostic adequacy (96%) and was overwhelmingly preferred by clinical care teams (nurses, anesthesiologists, and technologists) over conventional hardware. These results suggest that lightweight, blanket-like arrays can improve clinical workflow and patient safety without compromising image quality.

Introduction

In pediatric MRI, the hardware interface—specifically the radiofrequency (RF) receiver coil—is a frequent bottleneck for both image quality and patient comfort. Standard commercial coils are typically rigid and heavy. For neonates and small children, this weight can lead to respiratory constriction, necessitating the use of "bolsters" or foam supports to lift the coil off the patient. However, these supports create air gaps that increase the distance between the anatomy and the sensors, significantly degrading the Signal-to-Noise Ratio (SNR). Furthermore, the intimidating appearance of bulky hardware can increase anxiety in non-sedated children, leading to motion artifacts.

To address these issues, this research presents the development and clinical validation of a "form-fitting" receiver array designed to wrap around a patient like a blanket, maximizing the filling factor while minimizing physical burden.

Technical Innovation: AIR Technology

The core of this development is AIR technology, which represents a fundamental shift from traditional copper-strip array architectures.

Novel Conductors and Decoupling

Traditional coils rely on rigid copper loops and precise "critical overlap" to prevent mutual inductance (interference) between elements. The AIR technology utilizes a unique metal and dielectric multiple resonator conductor. This design significantly reduces weight while maintaining high-quality factors (Q). Crucially, it allows elements to "float" and overlap arbitrarily. This robustness ensures that the coil maintains signal decoupling (<-20dB) even when aggressively flexed or folded—a feat previously difficult to achieve with high-performance arrays.

Miniaturized Electronics

To further reduce weight and profile, the researchers integrated low-profile, noise-controlling preamplifiers directly at each loop. This co-location allows for superior control of magnetic and electric field coupling. The resulting 16-element array weighs just 480 grams (approximately 1 lb) excluding the interface cable, making it significantly lighter than the multi-kilogram conventional alternatives.

Clinical Implementation and Methodology

The study prospectively recruited 24 consecutive pediatric subjects (ranging from 23 days to 17 years of age) undergoing torso or extremity MRI. The flexible array was used in conjunction with either a high-density pediatric posterior array or a conventional posterior phased array.

Evaluation Metrics

  1. Diagnostic Adequacy: Images were reviewed by radiologists to ensure they met clinical standards.
  2. Care Team Preference: Surveys were administered to nurses, anesthesiologists, and radiologic technologists to compare the new array against standard cardiac and torso coils.
  3. Mechanical Performance: The coil's ability to maintain decoupling was tested in both flat and highly flexed states (e.g., wrapped around a 13 cm cylinder).

Results and Discussion

Clinical Acceptance

The feedback from the clinical care team was nearly unanimous. 100% of nurses and anesthesiologists involved in the cases preferred the flexible coil, citing the reduced risk of respiratory interference and ease of patient positioning. Technologists preferred the coil in 96% of cases, noting that the "blanket-like" nature of the device simplified the setup for complex pediatric cases.

Image Quality and Diagnostic Success

Of the 24 exams conducted, 23 (96%) were rated as diagnostically adequate. The single outlier was attributed to a coverage limitation for long bones rather than a failure of signal quality. Volunteer testing confirmed that the lightweight array provided SNR and image quality equivalent to conventional 16-channel cardiac arrays, proving that mechanical flexibility does not require a trade-off in performance.

Workflow Synergy

By eliminating the need for bolsters and towels to support the weight of the coil, the setup time was reduced. The ability to wrap the coil directly around a tiny limb or a 1.4 kg premature infant’s chest maximized the "filling factor," naturally boosting the SNR by placing the sensors as close to the source of the signal as possible.

Limitations and Future Directions

While the coil itself is exceptionally light, the interface cabling remains relatively heavy and bulky. Future iterations will focus on reducing the weight of the "tail" to make the device truly unobtrusive. Additionally, while the current biocompatible technical fabrics are functional, future designs could incorporate child-friendly aesthetics to further reduce scan-room anxiety.

Beyond pediatrics, this technology holds promise for fetal MRI, bariatric imaging, and the study of complex joints in adults where traditional rigid coils cannot provide adequate coverage.

Conclusion

The development of ultra-lightweight, highly flexible receiver arrays marks a significant step toward making MRI more "kid-friendly." By combining novel conductor materials with miniaturized electronics, this technology provides a safe, high-performance solution that meets the unique anatomical and physiological needs of pediatric patients while earning the strong preference of clinical providers.

Summary updated

Abstract

We develop a novel next generation light-weight highly flexible pediatric coil array, combine it with a high-density pediatric posterior array or conventional posterior phased array, and determine feasibility of pediatric clinical use. A highly flexible 16 element MRI receiver coil was constructed with low-profile noise controlling preamplifiers that minimized reactive and resistive coupling. Element decoupling was assessed in flat and highly flexed states. With IRB approval and informed consent and assent, 24 consecutive subjects undergoing torso or extremity MRI were prospectively recruited. Care team members were surveyed on preference for the coil versus conventional coils and diagnostic acceptability of the images was recorded. Confidence interval of proportion of diagnostic exams was calculated. The array without cable weighed 480 grams, demonstrated good flexibility while maintaining element decoupling. The coil was preferred by all nurses and anesthesiologists involved in the care of the patients. Technologists preferred the coil in 96 percent of cases, and 23 of 24 exams were diagnostically adequate with 85 percent confidence interval of 90-100 percent. Light-weight highly flexible coil arrays can be constructed that maintain element decoupling. Pediatric clinical image quality is likely to be diagnostic, with acceptance by members of the care team.
BibTeX
@misc{vasanawala2017development,
  author = {Shreyas Vasanawala and Robert Stormont and Scott A. Lindsay and Thomas Grafendorfer and Joseph Y. Cheng and John M. Pauly and Greig Scott and Jorge X. Guzman and Victor Taracila and Dan Chirayath and Fraser Robb},
  title = {{Development and Clinical Implementation of Next Generation Very Light Weight and Extremely Flexible Receiver Arrays for Pediatric MRI}},
  year = {2017},
  eprint = {1705.00224},
  archivePrefix = {arXiv},
  url = {https://arxiv.org/abs/1705.00224},
}