Determination of the Contents of Antioxidants and Their Degradation Products in Sodium Chloride Injection for Blood Transfusion

Bo Tao, Gang Wang, Zongning Yin, Xiaocong Pu, Yan Jiang, Luohong Zhang, Jie‐Fei Cheng, Yong Li, Jiayu ZhangJournal of Analytical Methods in Chemistry 2020 · DOI

TL;DR

This paper presents a novel HPLC method for simultaneously quantifying six common antioxidants and their degradation products in plastic infusion bags. Crucially, it demonstrates that only one specific degradation product, 1310, migrates into sodium chloride injection solutions at levels well below safety thresholds, ensuring drug safety.

What Problem Does This Paper Address?

Plastic infusion bags, often made from polyolefin polymers, are essential for administering intravenous fluids like sodium chloride injections for blood transfusions. To prevent these plastics from degrading and maintain their integrity, manufacturers add antioxidants. However, these antioxidants and their breakdown products can potentially migrate into the pharmaceutical solution, posing a risk to patient health. Existing analytical methods struggle to simultaneously detect both the parent antioxidants and their various degradation products, leaving a gap in comprehensive safety assessments. This study aims to fill that gap by developing and validating a method to accurately measure these compounds in both the packaging material and the final drug product.

Methods

The core of this research is the development and validation of a High-Performance Liquid Chromatography (HPLC) method. The authors utilized a Waters Symmetry RP18 column with a gradient elution system involving methanol, acetonitrile, and an acetic acid-water mixture. Detection was performed using UV at 277 nm, a wavelength chosen for its optimal absorption by the target compounds.

Key methodological aspects include:

  • Simultaneous Quantification: The method was designed to detect six specific antioxidants (BHT, 3114, 1010, 1330, 1076, and 168) and six of their degradation products (BHT-OH, BHT-COOH, 1310, BHT-CHO, BHT-Q, and DBP) in a single run.
  • Method Validation: The developed HPLC method underwent rigorous validation, assessing parameters like Limit of Detection (LOD), Limit of Quantification (LOQ), precision (intraday and interday), linearity, stability, and accuracy.
  • Sample Preparation: For packaging materials, a microwave-assisted extraction with dichloromethane followed by sonication in methanol was employed. For the sodium chloride injection samples, Solid-Phase Extraction (SPE) using a hydrophilic-lipophilic balance (HLB) cartridge was used to concentrate the analytes before HPLC analysis. This SPE step was crucial for improving detection sensitivity.

Datasets

Not discussed in the paper. The study focuses on developing and validating an analytical method using commercially available standards and real-world infusion bag samples.

Key Insights and Ideas

  • Degradation Products Matter: The study highlights that focusing solely on parent antioxidants is insufficient. Their degradation products can have different migration behaviors and potential toxicities.
  • Selective Migration: While several antioxidants and their degradation products were found in the plastic packaging, only one specific degradation product (1310) was detected migrating into the sodium chloride solution. This suggests that the water solubility and chemical properties of these compounds significantly influence their migration into aqueous solutions.
  • HPLC-UV as a Versatile Tool: The research demonstrates that HPLC with UV detection, specifically at 277 nm, is a powerful and cost-effective technique for simultaneously analyzing a range of antioxidants and their degradation products, overcoming limitations of other methods like GC-MS or LC-MS/MS for certain compounds.
  • Safety Assurance: The comprehensive analysis provides strong evidence that the migration of 1310 into sodium chloride injections for blood transfusion is well within acceptable safety limits, reassuring stakeholders about the safety of these widely used medical products.

Key Results

  • A validated HPLC method was established for the simultaneous determination of 12 analytes (6 antioxidants and 6 degradation products).
  • Method validation demonstrated excellent performance:
  • LOD: 0.011-0.151 µg/mL
    • LOQ: 0.031-0.393 µg/mL
    • Intraday Precision: 0.25%-3.17%
    • Interday Precision: 0.47%-3.48%
    • Linearity: r > 0.9994 across a range of 0.1-46.8 µg/mL
    • Stability: 0.35%-3.29%
    • Accuracy: 80.39%-104.31%
  • In the packaging materials, six antioxidants (BHT, 1010, 1330, 1076, 168) and two degradation products (1310, DBP) were detected.
  • In the migration study, only the degradation product 1310 was detected in the sodium chloride injection solution.
  • The maximum daily intake of 1310 from three bags of sodium chloride injection was found to be between 37.44 µg and 48.71 µg, which is significantly lower than the Permitted Daily Exposure (PDE) of 300 µg/day.

Limitations and Future Work

  • Limited Degradation Products: The study focused on a specific set of six degradation products. It's possible other, less common degradation products could exist and migrate.
  • Other Infusion Solutions: While the study focused on sodium chloride injection, future work could extend this methodology to other types of infusion solutions and different plastic formulations.
  • Long-Term Stability: While 12 months of stability testing were performed, even longer-term studies under various storage conditions could provide further reassurance.
  • Other Analytical Techniques: Although HPLC-UV proved effective, exploring complementary techniques like LC-MS/MS for even higher sensitivity or confirmation could be beneficial for future research.

Summary updated

Abstract

The infusion bag is mainly made up of polyolefin polymer. Antioxidants are usually added to these polymer materials in the production process to prevent the materials from aging and enhance the stability of the materials. Because of the potential harm of antioxidants to human body, it is necessary to limit the amount of antioxidants migrating to the pharmaceutical solutions. In the present study, we developed and validated the HPLC method for the simultaneous quantification of antioxidants and their degradation products migrating to sodium chloride solution for injection. A total of six antioxidants and six their degradation products were separated and simultaneously determined by using a Waters Symmetry RP18 column (250 × 4.6 mm, 5 <i>μ</i>m) and gradient elution of methanol/acetonitrile/acetic acid-water (1 : 99, v/v) at a flow rate of 1.0 mL/min. The detective wavelength was set at 277 nm, and the column temperature was maintained at 35°C. The method was validated in terms of limit of detection (LOD, 0.011-0.151 <i>μ</i>g/mL), limit of quantification (LOQ, 0.031-0.393 <i>μ</i>g/mL), intraday precision (0.25%-3.17%), interday precision (0.47%-3.48%), linearity (0.1-46.8 <i>μ</i>g/mL, <i>r</i> > 0.9994), stability (0.35%-3.29%), and accuracy (80.39%-104.31%). In the extraction experiment, antioxidants, BHT, 1010, 1330, 1076, and 168, and their degradation products, 1310 and DBP, were detected in the packaging materials. Only 1310 was detected in the migration experiment. The maximum daily dosage of sodium chloride for blood transfusion is three bags, and the content of 1310 in long-term testing samples is from 0 to 12 months ranging from 37.44 <i>μ</i>g/3 bags to 48.71 <i>μ</i>g/3 bags. The daily intake of 1310 did not exceed 48.71 <i>μ</i>g, which was much lower than its permitted daily exposure (PDE, 300 <i>μ</i>g/day). Therefore, the antioxidants and their degradation products migrating into the drug solution would not cause drug safety risks.
BibTeX
@article{tao2020determination,
  author = {Bo Tao and Gang Wang and Zongning Yin and Xiaocong Pu and Yan Jiang and Luohong Zhang and Jie‐Fei Cheng and Yong Li and Jiayu Zhang},
  title = {{Determination of the Contents of Antioxidants and Their Degradation Products in Sodium Chloride Injection for Blood Transfusion}},
  journal = {Journal of Analytical Methods in Chemistry},
  year = {2020},
  doi = {10.1155/2020/8869576},
}
Figure