Determination of Metal Elements in Urine by ICP-MS

Miao ZhangJournal of Environmental Health 2009

Abstract

Objective To establish a new method for the determination of metal elements in urine by ICP-MS. Methods Under the optimum operating conditions of ICP-MS instrument, cadmium, lead, copper, chromium, manganese, cobalt, arsenic, selenium, molybdenum, antimony, mercury and thallium in urine sample to be filtrated by 0.45 μm filter membrane before they were determined by ICP-MS using In, Sc, Y, Tb, Bi as the internal standards to compensate sample matrix effects. Results The linear ranges for 12 elements were 0-50 μg/L with a correlation coefficient for each element of more than 0.999 7, the detection limits were 0.056-0.57 μg/L, the recoveries were 72.5%-103.3%, the relative standard deviation was 1.5%-3.6%. The determined values for a certified ZK103-2, ZK018-2, ZK020-2 and ZK019-2 reference materials agreed to a permissible error range of the certified values. Conclusion This method is simple, quick, accurate, sensitive and is applicable to the determination of metal elements in urine.
BibTeX
@article{zhang2009determination,
  author = {Miao Zhang},
  title = {{Determination of Metal Elements in Urine by ICP-MS}},
  journal = {Journal of Environmental Health},
  year = {2009},
}

Loading related content...

Figure