Abstract: In this paper, the detection efficiency of a large area neutron sensitive microchannel plate (nMCP) detector has been evaluated. A $^6LiF/ZnS$ detector was employed as the benchmark detector; the Time-of-Flight (TOF) spectra of these two detectors were simultaneously measured, and the energy spectra were then deduced to calculate the detection efficiency curve of the nMCP detector. Tests show the detection efficiency at 25.3 meV thermal neutrons is 34% for this nMCP detector.
Introduction
Neutron radiography is a critical non-destructive testing (NDT) tool, offering unique sensitivity to light elements like hydrogen while penetrating heavy metals. High-resolution imaging requires detectors with exceptional spatial resolution, a role traditionally filled by microchannel plate (MCP) detectors. However, standard MCPs are insensitive to neutrons and must be doped with isotopes like Boron-10 or Gadolinium to capture neutrons and trigger secondary electron cascades.
A primary challenge in neutron imaging is detection efficiency. Low efficiency necessitates longer exposure times, which is problematic for neutron sources with limited flux. This study evaluates a 106 mm diameter MCP doped with 3 mol% of $natGd_2O_3$, aiming to demonstrate that gadolinium-doped designs can significantly improve efficiency over previous boron-doped iterations.
Experimental Methodology
The researchers utilized a comparative approach at the Compact Pulsed Hadron Source (CPHS) at Tsinghua University. By placing the nMCP detector and a standard $^6LiF/ZnS$ scintillator (with a known efficiency curve) in the beamline simultaneously, they could derive the nMCP's intrinsic efficiency.
Because the neutron source is pulsed, the "Time-of-Flight" (TOF)—the duration for a neutron to travel from the source to the detector—was measured. Since neutron speed is directly related to energy, this allowed the team to map detection efficiency across a broad energy spectrum. The analysis accounted for differences in distance, area, and electronic "dead time."
Results and Discussion
The experimental results confirmed that gadolinium doping provides a substantial boost in performance. While previous $^{10}B$-doped MCPs typically achieved approximately 16% efficiency, this nMCP detector reached 34% efficiency at the standard thermal neutron energy of 25.3 meV.
The efficiency curve demonstrates that the detector is most effective for cold and thermal neutrons, with efficiency decreasing as neutron energy increases. This behavior is consistent with the physics of neutron capture cross-sections. Additionally, the paper addresses "pulse smearing" caused by the 100 µs accelerator pulse width, which introduces uncertainty in energy calculations at higher energy levels.
Conclusion
The achievement of 34% efficiency brings high-resolution neutron imaging (previously demonstrated at 88 µm for this detector) into a more practical range for real-world industrial applications. While theoretical models suggest that optimized Gd-doped MCPs could reach up to 78% efficiency, this work represents a significant step forward. Future research will focus on optimizing plate geometry and conducting imaging trials on complex industrial components using low-intensity beams.
Summary updated