Design of Novel RF Pulse for Fetal MRI Refocusing Trains using Rank Factorization (SLfRank) to Reduce SAR and Improve Image Acquisition Efficiency
Lists
Only you see this
Abstract
Fetal MRI is inherently time-constrained due to unpredictable motion and comfort concerns. Single-shot spin-echo sequences address these concerns by rapidly acquiring images with desired tissue contrast. However, these sequences typically operate at specific absorption rate (SAR) limitations due to a train of refocusing pulses, which impose long TR times and reduce acquisition efficiency. This study leverages SLfRank, the use of ranked factorization to jointly solve for SLR bi-coupled spin parameters, to design refocusing profiles that reduce SAR while maintaining slice-profile performance. The proposed refocusing pulse reduced SAR by over 20% while producing similar image quality to the standard clinical sequence.
BibTeX
@inproceedings{diaz2023design,
author = {Sebastián Díaz and Yamin Arefeen and Borjan Gagoski and P. Ellen Grant and Elfar Adalsteinsson},
title = {{Design of Novel RF Pulse for Fetal MRI Refocusing Trains using Rank Factorization (SLfRank) to Reduce SAR and Improve Image Acquisition Efficiency}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2023},
doi = {10.58530/2022/2926},
}Loading related content...