Design of a Wireless Non-Contact Wearable System for Infants Using Adaptive Filter

Md Shaad Mahmud, Honggang Wang, Hua Fang2017

Abstract

Cardiovascular signals are fundamental physiological indicators for assessing health conditions. Continuous and long-term health monitoring of infants is crucial for the prevention and early prediction of illnesses. However, traditional monitoring methods in Neonatal Intensive Care Units (NICUs) often involve invasive adhesive patches that can damage fragile infant skin. This paper presents a wireless, non-contact wearable system designed to monitor infant ECG signals through clothing. The system integrates capacitive sensors with a 3-axis accelerometer and utilizes a Least Mean Square (LMS) adaptive filter to mitigate motion artifacts, providing a non-invasive alternative for continuous neonatal care.

Introduction

In the sensitive environment of a Neonatal Intensive Care Unit (NICU), continuous heart rate monitoring is essential. The current clinical standard relies on adhesive gel electrodes. For premature infants, whose skin is underdeveloped and extremely fragile, the repeated application and removal of these patches can lead to significant skin trauma and pain. Furthermore, the associated wiring restricts natural movement and complicates essential skin-to-skin contact between infants and their parents.

While non-contact capacitive sensors offer a solution to these physical constraints, they are highly susceptible to "motion artifacts"—electrical noise generated by the relative movement between the sensor and the body. This research addresses this challenge by developing a compact, wireless system that effectively filters out motion noise in real-time.

System Architecture

The developed prototype is a compact wearable device measuring 51mm x 16mm x 6mm, designed for integration into an infant's diaper. The hardware stack includes:

  • Non-Contact Sensors: Two EPIC (Electric Potential Integrated Circuit) capacitive electrodes capable of measuring ECG signals through layers of clothing or insulators.
  • Analog Front End (AFE): An AD8232 heart rate monitor front-end chip provides high-gain amplification and initial analog filtering.
  • Motion Tracking: An ADXL 345 3-axis accelerometer samples physical movement at 80 Hz.
  • Processing and Communication: A "Simblee" Bluetooth Low Energy (BLE) module manages data acquisition, real-time processing, and wireless transmission to a mobile application.
The hardware prototype (51mm x 16mm) and the PS25255 non-contact sensor compared to a coin.

Adaptive Filtering for Noise Cancellation

The primary innovation of this system lies in its approach to noise cancellation. Because motion artifacts are unpredictable in both amplitude and phase, traditional fixed-coefficient filters are ineffective. Instead, the team implemented a Least Mean Square (LMS) adaptive filter.

The algorithm utilizes the 3-axis accelerometer data as a "noise reference." By calculating the sum of the X, Y, and Z axes, the system creates a comprehensive profile of physical movement. The LMS filter then dynamically adjusts its parameters to identify the correlation between the accelerometer data and the noise in the ECG signal, mathematically subtracting the artifacts to recover the underlying cardiac waveform.

Experimental Results

The system was evaluated using data from adult subjects to simulate the placement and movement conditions expected in a neonatal setting. The results demonstrated a significant improvement in signal clarity.

Comparison of the raw noisy ECG signal (top) and the cleaned signal after processing through the LMS adaptive filter (bottom).

Key findings from the testing phase include:

  • Optimal Filter Parameters: Through 72 independent trials, the researchers determined that a step-size (μ) of 0.016 and a filter order (M) between 16 and 24 yielded the best performance, balancing noise reduction with computational efficiency.
  • Robustness: The 3-axis fusion approach ensured the filter remained effective regardless of the sensor's orientation or the direction of movement.
  • Efficiency: The system is designed for low power consumption, with the instrumentation amplifier drawing only 180 μA, supporting long-term monitoring on a single battery charge.

Conclusion and Future Directions

This research demonstrates the feasibility of a non-contact, wireless ECG monitoring system that can withstand the motion artifacts typical of wearable applications. By replacing rigid adhesive patches with a diaper-integrated sensor node, this technology has the potential to significantly improve the quality of care for infants in NICUs. Future work will focus on transitioning the design to flexible PCBs for enhanced comfort and conducting clinical trials with neonatal subjects to validate the system's performance in a real-world medical environment.

Summary updated

Abstract

Cardiovascular signal is a fundamental physiological sign to assess health condition. Continuous and long term health monitoring of infants would be helpful to prevent and predict illness. However, there is no noninvasive method to monitor their health. The developed system consists of a non-contact
BibTeX
@misc{mahmud2017design,
  author = {Md Shaad Mahmud and Honggang Wang and Hua Fang},
  title = {{Design of a Wireless Non-Contact Wearable System for Infants Using Adaptive Filter}},
  year = {2017},
  doi = {10.4108/eai.13-7-2017.2270652},
}
Figure