Design and Simulation of a high-resolution and high-sensitivity BrainPET insert for 7T MRI
Lists
Only you see this
Abstract
Ziel/Aim Within the Helmholtz Validation Fund Project "Next generation BrainPET scanner for 7T MRI", we aim at building a UHF-MRI compatible, BrainPET insert prototype for dedicated human neuroimaging. The BrainPET 7T insert was designed to achieve a sensitivity at the isocentre of » 12 %, a homogeneous spatial image resolution between 1.5 and 2.0 mm over the whole human brain, a coincidence timing resolution £ 500 ps, and an axial field of view of » 24 cm.
BibTeX
@inproceedings{lerche2020design,
author = {Christoph Lerche and Mirjam Lenz and Wenli Bi and J. Scheins and Lutz Tellmann and CH Choi and ER Kops and Jörg Felder and D. Arutinov and Sebastian H. Völkel and R.D. Heil and J Collienne and Dirk Grunwald and Bjoern Weissler and Fabian Müller and David Schug and Volkmar Schulz and JL Lefaucheur and Zhen Chen and Gary F. Egan and N. Jon Shah},
title = {{Design and Simulation of a high-resolution and high-sensitivity BrainPET insert for 7T MRI}},
booktitle = {Nuklearmedizin},
year = {2020},
doi = {10.1055/s-0040-1708248},
}Loading related content...