Densimeter calibration method versus temperature and pressure
Lists
Only you see this
Abstract
This article reports a calibration procedure for the commercial Anton Paar DMA 512 densimeter usable at pressures up to 40 MPa. The method is based on water as a reference substance and requires knowledge of the characteristics of the cell as functions of temperature when subjected to a vacuum lower than 100 Pa. Comparis on with data from the literature indicates a relative accuracy of 0.1% where the classical calibration method (i.e. assuming no dependence of the densimeter performance upon pressure) led to a relative accuracy of 0.8% for the same substances.
BibTeX
@inproceedings{lagourette1992densimeter,
author = {B. Lagourette and Christian Boned and H. Saint‐Guirons and P. Xans and Han Zhou},
title = {{Densimeter calibration method versus temperature and pressure}},
booktitle = {Measurement},
year = {1992},
doi = {10.1088/0957-0233/3/8/002},
}Loading related content...