Intra-operative brain shift is a well-known phenomenon that describes non-rigid deformation of brain tissues due to gravity and loss of cerebrospinal fluid among other phenomena. This has a negative influence on surgical outcome that is often based on pre-operative planning where the brain shift is not considered. We present a novel brain-shift aware Augmented Reality method to align pre-operative 3D data onto the deformed brain surface viewed through a surgical microscope. We formulate our non-rigid registration as a Shape-from-Template problem. A pre-operative 3D wire-like deformable model is registered onto a single 2D image of the cortical vessels, which is automatically segmented. This 3D/2D registration drives the underlying brain structures, such as tumors, and compensates for the brain shift in sub-cortical regions. We evaluated our approach on simulated and real data composed of 6 patients. It achieved good quantitative and qualitative results making it suitable for neurosurgical guidance.
BibTeX
@inproceedings{haouchine2020deformation,
author = {Nazim Haouchine and Parikshit Juvekar and William M. Wells III and Stephane Cotin and Alexandra J. Golby and Sarah F. Frisken},
title = {{Deformation Aware Augmented Reality for Craniotomy Using 3D/2D Non-rigid Registration of Cortical Vessels}},
booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
year = {2020},
url = {https://openreview.net/forum?id=HUYVoUcbjE},