Deep learning to assess laryngoscope insertion depth during neonatal intubation with video laryngoscopy

Abrar Majeedi, Patrick Peebles, Yin Li, Ryan M. McAdamsJournal of Perinatology 2025 · DOI

Abstract

AI-enabled video laryngoscopy (VL) may support blade-insertion depth assessment and training. Given the rarity of deep events, conclusions about deep insertion and clinical impact are limited. Future multi-site studies should focus on clinical integration and assessing outcomes.

Introduction

Endotracheal intubation is a critical yet technically demanding procedure in neonatal intensive care, particularly for preterm infants with unique anatomical challenges. First-pass success rates for trainees are often below 50%, and repeated attempts are associated with significant risks, including bradycardia and intraventricular hemorrhage.

A primary determinant of success is the laryngoscope blade insertion depth. Incorrect depth—either too shallow (obscuring the glottis) or too deep (entering the esophagus)—is a common cause of failure. While video laryngoscopy (VL) provides improved visualization, clinicians currently lack real-time, automated guidance to optimize blade positioning. This paper presents a deep learning system designed to provide such feedback.

Methods

The researchers developed a spatiotemporal deep learning pipeline to classify insertion depth into three categories: "Shallow," "Glottic Zone," and "Deep."

Model Architecture and Processing

The system utilizes an Inflated 3D Convolutional Network (I3D) backbone to capture both spatial features and temporal dependencies across video frames. The model processes 8-frame clips at 10 Hz, providing 0.8 seconds of temporal context for each prediction. To ensure stable feedback and minimize "flickering" in the user interface, a five-frame sliding window with majority voting was implemented as a post-processing step.

Overview of the spatiotemporal deep learning pipeline using an I3D backbone and sliding window for real-time classification.

Dataset and Labeling

The study leveraged 132 neonatal VL recordings from two neonatal intensive care units (NICUs), totaling 298,955 annotated frames. Expert neonatologists labeled the frames based on anatomical landmarks:

  • Shallow (24.9%): Glottic structures not yet visible.
  • Glottic Zone (72.4%): Any view where glottic structures are visible.
  • Deep (2.7%): Landmarks bypassed, typically entering the esophagus.

Results

The model demonstrated high performance in identifying the primary zones of interest. It achieved an F1 score of 0.894 for the "Glottic Zone" and 0.718 for "Shallow" insertions. The temporal smoothing mechanism effectively reduced transient mispredictions, providing a more reliable signal for clinical use.

Precision-Recall curves for the Shallow and Glottic Zone classification tasks, showing robust performance in these categories.

However, the model struggled with the "Deep" insertion category, yielding an F1 score of only 0.034. This low performance is attributed to a significant data imbalance; expert clinicians rarely over-insert the blade, resulting in very few training examples for this specific error.

Human-Centered Design

A survey of 31 clinicians (including neonatologists, fellows, and nurse practitioners) was conducted to determine optimal feedback delivery. Participants strongly preferred minimalistic visual prompts that appear only when adjustments are necessary. Auditory and haptic feedback were less favored, likely due to the high-stress, alarm-saturated environment of the NICU.

Discussion and Future Work

This study demonstrates the feasibility of using deep learning to provide real-time feedback on laryngoscope insertion depth. By defining a broad "Glottic Zone," the system identifies the "neighborhood" of success, supporting clinicians during the approach. The system was also successfully integrated with a previous "glottic opening detection" model, creating a multi-tiered AI assistant.

Future research must address the "Deep" insertion detection gap, potentially through simulated data. Additionally, while the model can accurately classify depth on the Karl Storz C-MAC platform, further validation is needed across different hardware. The ultimate goal is to conduct prospective clinical trials to evaluate whether this AI assistance improves first-pass success rates and neonatal outcomes.

Summary updated

Abstract

AI-enabled VL may support blade-insertion depth assessment and training. Given the rarity of deep events, conclusions about deep insertion and clinical impact are limited. Future multi-site studies should focus on clinical integration and assessing outcomes.
BibTeX
@article{majeedi2025deep,
  author = {Abrar Majeedi and Patrick Peebles and Yin Li and Ryan M. McAdams},
  title = {{Deep learning to assess laryngoscope insertion depth during neonatal intubation with video laryngoscopy}},
  journal = {Journal of Perinatology},
  year = {2025},
  doi = {10.1038/s41372-025-02457-0},
}
Figure