ConRef: Auditing Reference Reliability in 2D/3D Registration Benchmarks

Jianfeng Yue, Feng Li, Tianyu Song, Julius Watrinet, Ulrich Eck, Nassir NavabarXiv 2026

Imagine you’re a high-performance watchmaker. You’ve spent years perfecting a movement that is accurate to within a millisecond. But when you go to test it, the "official" atomic clock you’re using as a reference is itself drifting by a few seconds every day. How do you know if your watch is actually good, or if you’re just measuring the errors of the clock?

This is the exact dilemma facing researchers in 2D/3D medical image registration, and it's the problem the authors of ConRef set out to solve.

TL;DR

As medical AI gets better at aligning 3D anatomy (like CT scans) with 2D X-rays, we’ve hit a "noise floor" where our benchmark datasets aren't as accurate as the algorithms we’re testing. ConRef is a new framework that uses multi-view geometry to "audit" these benchmarks, providing a more reliable "consensus" reference and flagging hidden errors that could lead researchers to the wrong conclusions.

What Problem Does This Paper Address?

In orthopedic surgery, "registration" is the process of mapping a patient's preoperative 3D CT scan onto the 2D X-rays taken during surgery. It’s like trying to perfectly orient a 3D bone model so its shadow matches the X-ray on the screen.

For years, we’ve relied on public benchmarks to tell us which algorithms are best. These benchmarks come with "reference poses"—the supposed ground truth. However, these references are often created using markers (like metal pins) or manual alignment, both of which have their own margins of error.

Reference uncertainty can distort measured error as methods improve. At submillimeter scales, the reference error can actually be larger than the registration error itself.

As shown in Figure 1a, when our algorithms were "Past" (10mm error), a 1mm reference error didn't matter. But today, in the "Present," our algorithms are chasing submillimeter accuracy. If the reference is off by 1mm, an algorithm with 0.5mm true error might be measured as having 1.5mm error. We are effectively grading a genius student with a flawed answer key.

Methods

The authors propose Consensus Reference (ConRef). Instead of trusting a single provided reference, ConRef looks at all available X-ray views of a single patient/phantom simultaneously.

  1. Multi-view Optimization: It finds a single 3D pose that best fits all the X-ray projections at once. By jointly exploiting multiple observations, it creates a "consensus" that is statistically more robust than any single-view estimate.
  2. Uncertainty Quantification: Using a technique called a "delete-k jackknife," the system estimates how much the consensus would change if some X-rays were missing. This gives a 95% confidence radius for the reference itself.
  3. Anomaly Detection: It performs "consistency checks." If one specific X-ray view wildly disagrees with the consensus of the others, it gets flagged as a potential outlier (e.g., due to a bad camera calibration or a moved patient).

Datasets

The study evaluated ConRef on three distinct datasets:

  • DeepFluoro: A widely used benchmark of cadaveric pelvises.
  • Femur: A dataset of cadaveric distal femurs using K-wire markers for reference.
  • BeadPhant (New): The authors introduced this high-precision phantom dataset. It uses synthetic bones fixed to acrylic plates with steel beads. Because the beads' positions are known with extreme precision (~0.1mm), it provides a "gold standard" to validate whether ConRef actually works.
The BeadPhant dataset uses steel beads on an acrylic plate to establish a highly precise ground truth.

Key Insights and Ideas

  • The "Success" Flip: The most startling insight is that the choice of reference can completely change the scientific conclusion of a study. On one femur dataset, using the original reference made it look like no registration method could achieve submillimeter accuracy. Switching to the ConRef or anatomical reference showed that the methods were actually succeeding.
  • Redundancy is an Auditor: You don't need a perfect "ground truth" to find errors; you just need enough overlapping data. If ten X-rays agree on a pose and the eleventh is 5mm off, you've found a bug in the dataset, not the algorithm.
  • Markers Aren't Perfect: Even "gold standard" markers like K-wires can shift slightly between the CT scan and the X-ray room, introducing systematic bias that ConRef can detect.

Key Results

  • Validation: On the BeadPhant dataset, ConRef stayed within 0.08–0.12 mm of the bead-based reference, proving it is highly accurate.
  • Auditing Benchmarks: ConRef found that the Femur dataset had discrepancies as large as 2.06 mm compared to the original references.
  • Outlier Hunting: It flagged 22 X-rays in the DeepFluoro dataset as potential outliers. Interestingly, three of these were already known to be problematic by the original dataset creators, but ConRef found 19 more.
  • Impact on Evaluation: As seen in Figure 6, when evaluating a contour-based registration method, the "Measured Error" (mTRE) drops significantly when moving from the original reference to ConRef, revealing that the algorithms were performing much better than previously thought.
Comparison of registration errors against different references. Note how the median error (black bar) drops below the 1mm threshold when using ConRef or AnatRef compared to the original reference.

Limitations and Future Work

While ConRef is a powerful auditor, it isn't magic. Its accuracy still depends on the angular coverage of the X-rays; if all X-rays are taken from roughly the same direction, the "depth" of the 3D object remains uncertain. Furthermore, if there are systematic errors shared across all views—like a consistently miscalibrated X-ray source—ConRef might agree on a "consensus" that is still wrong.

The authors suggest that future benchmarks should include these kinds of consensus audits as a standard part of their release, ensuring that as our algorithms get sharper, our yardsticks do too.

Summary updated

Abstract

Purpose: Accurate 2D/3D registration is crucial for image-guided orthopedic interventions, enabling the alignment of preoperative 3D anatomy with intraoperative X-ray observations for surgical navigation. Public benchmarks have facilitated the development and evaluation of increasingly accurate registration methods. However, as registration errors approach the uncertainty of benchmark reference poses, measured performance may be limited by reference inaccuracies rather than algorithmic errors alone. Methods: We propose Consensus Reference (ConRef), a multi-view framework for auditing reference poses in existing X-ray/computed tomography (CT) registration benchmarks. By jointly exploiting multiple X-ray observations, ConRef estimates an additional reference pose, quantifies its uncertainty, and identifies potential X-ray outliers. We validate the framework using independent anatomical evidence and a newly acquired X-ray/cone-beam CT (CBCT) phantom dataset with bead-based reference poses. Results: On our BeadPhant dataset, ConRef differed from independent bead-based references by 0.08-0.12 mm, while discrepancies with original references reached 2.06 mm on the Femur dataset. It also flagged 22 DeepFluoro X-rays as potential outliers. On Femur F5, reference choice changed the submillimeter success conclusions of the same registrations. Conclusion: Our findings demonstrate that reference discrepancies can influence the evaluation of high-accuracy 2D/3D registration methods. By providing an additional, independently validated reference and identifying potential inconsistencies, ConRef enables systematic auditing of existing benchmarks and supports more reliable evaluation of registration error.