Conducting interslice stimulation for concurrent TMS-fMRI
Abstract
A 10 Hz TMS interslice protocol is possible with minimal data contimination. A stimulation frequency faster than 10 Hz would require a shorter gap or shorter slice acquisition times. Further, stimulator intensity, slice orientation, and the number of TMS pulses affected data quality and are important considerations for researchers when setting up their own protocol.
BibTeX
@article{jackson2025conducting,
author = {J.B. Jackson and Catriona L. Scrivener and Marta Correia and Marius Mada and Alexandra Woolgar},
title = {{Conducting interslice stimulation for concurrent TMS-fMRI}},
journal = {Journal of Neuroscience Methods},
year = {2025},
doi = {10.1016/j.jneumeth.2025.110513},
}Ask Althea to put this paper to the test.