Concordance between side-stream end-tidal carbon dioxide and arterial carbon dioxide partial pressure in respiratory service setting.
Lists
Only you see this
Abstract
End-tidal carbon dioxide partial pressure did not show significant correlation or concordance with arterial carbon dioxide partial pressure, especially when supplemental oxygen was used. End-tidal carbon dioxide partial pressure currently cannot replace arterial blood gas measurement as a tool for monitoring arterial carbon dioxide partial pressure. Possible reasons for the discrepancy with previous studies include small sample size in previous studies, lack of research facilities in service settings, and publication bias against negative studies.