Comparison of respiratory navigator techniques for interleaved high-resolution coronary vessel wall imaging

Mehmet Akçakaya, Markus Henningsson, Reza Nezafat, René BotnarJournal of Cardiovascular Magnetic Resonance 2013 · DOI

Introduction

Imaging the coronary arteries presents a significant technical challenge in magnetic resonance imaging (MRI). These vessels are small, tortuous, and subject to constant motion from both the cardiac cycle and the respiratory cycle. While standard coronary angiography focuses on the vessel lumen, visualizing the vessel wall itself is essential for identifying early-stage atherosclerosis and characterizing plaque morphology.

To achieve high-resolution vessel wall imaging, researchers often employ "interleaved" sequences. These techniques, such as interleaved T2-prepared (i-T2prep) sequences, acquire multiple volumes with different contrast properties (e.g., with and without a preparation pulse). By processing these volumes together—often through subtraction—the signal from blood can be suppressed, highlighting the vessel wall. However, the success of this approach depends heavily on precise voxel registration between interleaves, making accurate respiratory motion compensation critical.

Methods

This study evaluated two distinct respiratory gating strategies for high-resolution coronary vessel wall MRI: synchronous and asynchronous gating. The research was conducted on seven healthy volunteers (average age 29.6 ± 12.6 years) using a 1.5T MRI scanner. The primary anatomical target for all scans was the Right Coronary Artery (RCA).

MRI scan identifying the Right Coronary Artery (RCA), the primary anatomical target for the vessel wall imaging study.

Both strategies utilized a 5mm respiratory gating window, but differed in how they accepted data:

  1. Synchronous Gating: Data is only accepted if the respiratory position for both interleaves in a pair (the T2prep+ and T2prep- heartbeats) falls within the 5mm window.
  2. Asynchronous Gating: Each interleaf is accepted independently as long as its specific heartbeat falls within the 5mm window, regardless of the position of its paired interleaf.

The resulting images were processed via weighted subtraction to generate vessel wall depictions. The performance of each method was quantified using two primary metrics: total acquisition time and vessel sharpness.

Key Insights: The Speed-Quality Trade-off

The core finding of this study is that "voxel-by-voxel" processing—whether for subtraction-based vessel wall imaging or quantitative T1 mapping—is far more sensitive to motion than standard anatomical imaging. In traditional coronary MRI, a 5mm gating window is often considered a reasonable compromise between scan time and motion blur. However, for interleaved sequences, the absolute position of the diaphragm is less important than the relative position between the interleaves.

Even a small 5mm window allows for enough internal motion to degrade the registration between the T2prep+ and T2prep- frames. Asynchronous gating is much more "efficient" at collecting data because it doesn't wait for matching pairs, but it is "inefficient" at maintaining the precise registration required for high-fidelity subtraction. This results in the observed decrease in vessel sharpness.

Key Results

The study revealed a significant trade-off between imaging speed and image quality:

  • Acquisition Efficiency: Asynchronous gating was substantially faster, with an average scan time of 9:29 minutes. In contrast, the synchronous approach required an average of 18:18 minutes (p=0.005). This nearly two-fold increase in speed makes asynchronous gating highly attractive for clinical workflows.
  • Image Sharpness: Despite the speed advantages of the asynchronous method, synchronous gating yielded superior image quality. The vessel wall sharpness was significantly higher in the synchronous group (0.15 ± 0.02 vs. 0.13 ± 0.02, p=0.03).
Comparison of coronary vessel wall imaging results. The subtraction process highlights the vessel wall, but requires precise registration between interleaved frames.

Discussion and Conclusion

The results underscore that for processing techniques like vessel wall subtraction, simple gating is often insufficient. While asynchronous gating offers a significant reduction in scan time, it does so at the cost of the fine structural detail required to accurately delineate the coronary vessel wall. Synchronous gating ensures that the two images being subtracted are acquired at nearly identical respiratory phases, maintaining the registration necessary for high-resolution detail.

Future research may focus on "intra-window" motion correction algorithms. Such techniques could potentially allow for the speed of asynchronous acquisition while computationally re-aligning data to match the sharpness of synchronous gating. Additionally, further testing in patients with irregular breathing patterns is necessary to determine the clinical robustness of these strategies.

Summary updated

Abstract

Background Three-dimensional high-resolution PSIR LGE [1], flowindependent vessel wall imaging [2], or T1 mapping [3] are all implemented with interleaved sequences, where the same volume is imaged multiple times with different inherent contrast. These volumes are then processed together to generate additional information, e.g. recovering the sign of magnetization, depicting vessel walls via subtraction, or calculating T1 maps via curve-fitting. Such techniques rely on correct registration of the voxels, hence on accurate motion compensation. In this study, we sought to explore the effects of synchronous and asynchronous navigator gating and tracking on respiratory motion compensation in high-resolution coronary vessel wall imaging with an interleaved T2prep sequence (i-T2prep).
BibTeX
@article{akcakaya2013comparison,
  author = {Mehmet Akçakaya and Markus Henningsson and Reza Nezafat and René Botnar},
  title = {{Comparison of respiratory navigator techniques for interleaved high-resolution coronary vessel wall imaging}},
  journal = {Journal of Cardiovascular Magnetic Resonance},
  year = {2013},
  doi = {10.1186/1532-429x-15-s1-e20},
}
Figure