Comparison of Aqueous Outflow Facility Measurement by Pneumatonography and Digital Schiøtz Tonography

Arash Kazemi, Jay W. McLaren, Shuai-Chun Lin, Carol B. Toris, Vikas Gulati, Sayoko E. Moroi, Arthur J. SitInvestigative 2017

Abstract

Outflow facilities measured by pneumatonography were correlated with those measured by Schiøtz tonography, but Schiøtz tonography reported approximately 10% to 20% higher facilities when using the standard method. When ocular rigidity was determined for each eye, differences were smaller. Measurements from these devices cannot be compared directly.
BibTeX
@inproceedings{kazemi2017comparison,
  author = {Arash Kazemi and Jay W. McLaren and Shuai-Chun Lin and Carol B. Toris and Vikas Gulati and Sayoko E. Moroi and Arthur J. Sit},
  title = {{Comparison of Aqueous Outflow Facility Measurement by Pneumatonography and Digital Schiøtz Tonography}},
  booktitle = {Investigative},
  year = {2017},
  doi = {10.1167/iovs.16-20754},
}

Loading related content...

Figure