Comparison of Aqueous Outflow Facility Measurement by Pneumatonography and Digital Schiøtz Tonography
Abstract
Outflow facilities measured by pneumatonography were correlated with those measured by Schiøtz tonography, but Schiøtz tonography reported approximately 10% to 20% higher facilities when using the standard method. When ocular rigidity was determined for each eye, differences were smaller. Measurements from these devices cannot be compared directly.