Coded-aperture DUV spectrometer for stand-off Raman spectroscopy
Lists
Only you see this
Abstract
We have designed and constructed a coded aperture spectrometer for use in the deep UV range. The Czerny-Turner design provides sufficient spectral resolution to observe Raman scattering features, while the use of a coded aperture provides a greatly improved light collection efficiency for scattering sources. The resulting instrument is capable of analyzing Raman spectra from samples at a 1 meter viewing distance. We provide an overview of the system, its optical design, and some preliminary measurements.
BibTeX
@inproceedings{hagen2009coded,
author = {Nathan Hagen and David J. Brady},
title = {{Coded-aperture DUV spectrometer for stand-off Raman spectroscopy}},
booktitle = {Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE},
year = {2009},
doi = {10.1117/12.818830},
}Loading related content...