CMOS Image Sensor Design and Image Processing Algorithm Implementation for Total Hip Arthroplasty Surgery

Syed Mudassir Hussain, Fasih Ud Din Farrukh, Shaojie Su, Zhihua Wang, Hong ChenIEEE Trans. Biomed. Circuits Syst 2019

Abstract

Total Hip Arthroplasty (THA) is a critical procedure for patients with damaged hip joints, yet it faces a significant failure rate due to improper prosthetic placement. This paper presents a specialized visual-aided system featuring a custom-designed CMOS image sensor and an FPGA-based processing pipeline. The 200 × 200 resolution sensor is optimized for low power and small form factor (3.5 mm x 3.5 mm) to fit within a femoral head prosthesis. Coupled with a novel blood-resistant pattern recognition algorithm, the system achieves a 99% detection rate and a 70% reduction in power consumption compared to previous iterations, offering a robust solution for real-time surgical navigation.

Introduction: The Challenge of Precision in THA

Total Hip Arthroplasty (THA) is increasingly common as populations age, yet approximately 10% of these surgeries fail. These failures are often attributed to the misalignment of prosthetic components outside the "safe zone," leading to dislocation or restricted mobility. While navigation systems exist, they are often too bulky for integration into surgical tools or consume excessive power for wireless operation. Furthermore, the surgical environment is typically obscured by blood and uneven lighting, which hampers standard optical tracking methods.

System Architecture and Custom Hardware

To address these challenges, the researchers developed an integrated "smart" prosthesis system. The core of the system is a custom-designed CMOS image sensor array embedded within the femoral head. This sensor tracks customized patterns placed within the acetabular cup (pelvis) to estimate the prosthetic's pose in real-time.

System overview of the visual-aided THA navigation system, showing the sensor array integrated into the femoral head prosthesis and the reference patterns in the pelvis.

The hardware design prioritizes miniaturization and efficiency. By moving away from commercial camera modules, the team achieved a chip footprint of just 3.5 mm x 3.5 mm. The sensor utilizes a 4-transistor (4T) pixel architecture, which is specifically tuned to handle the low-light variations encountered inside the joint.

Circuit schematic of the 4-transistor (4T) pixel architecture and the physical layout of the 200 x 200 resolution sensor chip.

Blood-Resistant Image Processing Pipeline

A major innovation of this work is the image processing pipeline implemented on a Xilinx Virtex 7 FPGA. Processing data pixel-by-pixel locally on the FPGA significantly reduces the power required for wireless data transmission. The pipeline includes:

  1. Brightness Control: Compensates for uneven illumination within the surgical site.
  2. Red Color Segmentation: A specialized step that treats blood as a filter rather than noise. By isolating the red channel, the system can more effectively segment the black-and-white reference patterns from the blood-covered background.
  3. Morphological Operations: Median filtering and "opening" techniques clean the segmented image, allowing for the accurate extraction of 9-bit binary IDs from the markers.

Key Results and Performance

The proposed system demonstrates significant improvements over existing methods:

  • Power Efficiency: The total system power consumption is 213 mW, representing a 70% reduction from previous designs. This is largely due to the custom sensor and local FPGA processing.
  • Accuracy: The pattern detection rate reached 99%, outperforming the standard "top-hat" algorithm by 5%.
  • Sensor Linearity: Simulations showed a linear response across input currents from 2 pA to 100 pA, resulting in a measurable voltage range of 2 mV to 855 mV.

Conclusion and Future Directions

This work successfully integrates a custom CMOS sensor and a robust processing pipeline into a form factor suitable for THA navigation. Future work will focus on further increasing computational speed through parallel block processing and moving toward a full System-on-Chip (SoC) integration. While initial tests using simulated blood (red ink) are promising, clinical validation in real surgical environments remains the next critical step.

Summary updated

Abstract

As the average age of the population goes higher, the people undergo hip arthroplasty surgery to replace their stiff and painful damaged hip joints. To reduce the risk factor after the surgery we developed a visual aided system in previous works. In order to solve the integration problems of commercial camera with other sensors into the femoral head and to minimize the area and power consumption, in this paper we propose a CMOS image sensor of resolution 200 × 200 specially designed for the application in which each individual pixel measures around 15 μm × 15 μm in size and the image sensor chip size measures about 3.5 mm × 3.5 mm. The proposed sensor is simulated with the input current variations from 2 pA to 100 pA for the individual pixels and the corresponding measurements for each pixel range from 2 mV to 855 mV. Besides, we put forward a new method of pattern detection and recognition in the blood-covered situation, which provides an accurate segmentation of patterns from the blood. All the detected patterns are recognized by generating its right 9-bit binary ID required for the pose estimation calculation. Furthermore, to reduce system power consumption, we implement algorithms on FPGA to process the image data pixel by pixel and transmit it directly to the computer for post-processing. Experimental results show that the pattern detection rate goes as high as 99%, which is 5% better in accuracy compared to the top hat algorithm. The power consumption of the system is 213 mW, which is a 70% decrease compared to our previous work.
BibTeX
@article{hussain2019cmos,
  author = {Syed Mudassir Hussain and Fasih Ud Din Farrukh and Shaojie Su and Zhihua Wang and Hong Chen},
  title = {{CMOS Image Sensor Design and Image Processing Algorithm Implementation for Total Hip Arthroplasty Surgery}},
  journal = {IEEE Trans. Biomed. Circuits Syst.},
  year = {2019},
  url = {https://openreview.net/forum?id=sgFQMi5Bix},
}
Figure