Characterizing On-Road Vibrations in Ontario's Neonatal Patient Transport
Abstract
Vibration and noise exposure during emergency neonatal ground and air transport are recognized as hazards for the neonate; however, no standardized test environment exists to reliably quantify such exposure with high repeatability. An on-road test was performed to provide a quantitative measurement of on-road vibrations to be used in the development of the standardized test environment. This paper details the ambulance on-road test instrumentation, analysis, and results. On-road findings indicate that neonates experience vibrations that exceed adult standards; that the power spectral density of the acceleration at the neonate contains peaks (at 2 Hz and 9 Hz) in the frequency range humans are most sensitive (1-20 Hz); and that the damped natural frequency of the patient transport system considered is 9.5 Hz.
BibTeX
@inproceedings{darwaish2021characterizing,
author = {F. Darwaish and Stephanie Redpath and Kim Greenwood and Cheryl Aubertin and A. A. M. Ibey and Adrian D. C. Chan and James R. Green and Robert G. Langlois},
title = {{Characterizing On-Road Vibrations in Ontario's Neonatal Patient Transport}},
booktitle = {MeMeA},
year = {2021},
url = {https://openreview.net/forum?id=LSbbrsu4-o},
}Ask Althea to put this paper to the test.