Characterisation of a Silicon Photomultiplier Based Oncological Brachytherapy Fibre Dosimeter

Massimo Caccia, Agnese Giaz, Marco Galoppo, Romualdo Santoro, Micheal Martyn, Carla Bianchi, Raffaele Novario, Peter Woulfe, Sinead O'Keeffe2024

TL;DR

This paper introduces and characterizes a novel fiber-optic dosimeter system for real-time, in-vivo dose monitoring during brachytherapy. By integrating silicon photomultipliers (SiPMs) with scintillating fiber tips, the system aims to provide crucial source localization and dose verification, addressing a critical need for improved patient safety and treatment efficacy in cancer therapy.

What Problem Does This Paper Address?

Brachytherapy, a cancer treatment involving the implantation of radioactive sources, relies heavily on accurate dose delivery. However, current methods for verifying this dose are often indirect (e.g., pre- and post-treatment imaging) and lack real-time, in-vivo capabilities. This can lead to undetected errors, potentially causing under- or over-dosing of the target tumor and surrounding healthy tissues, resulting in adverse side effects and compromised treatment outcomes. Existing real-time dosimetry solutions are limited in their spatial coverage and accuracy. Furthermore, there's a lack of commercial systems for real-time monitoring of radioactive source location. This paper tackles this gap by developing and characterizing a multi-point, fiber-optic based dosimeter system that offers both real-time dose measurement and source localization.

Methods

The core of the ORIGIN system is a silicon photomultiplier (SiPM) coupled to a polymer optical fiber with a scintillating tip. When radiation interacts with the scintillator, it emits light, which is then detected by the SiPM. The system was designed to operate in counting mode, leveraging the single-photon sensitivity of SiPMs.

Key methodological aspects include:

  • SiPM Selection: A crucial step involved characterizing and selecting the optimal SiPMs from different manufacturers (KETEK and HAMAMATSU) based on parameters like Photon Detection Efficiency (PDE), sensitive area, Dark Count Rate (DCR), and Optical Cross-Talk (OCT).
  • Scintillator Choice: Two types of scintillators, Y₂O₃:Eu + YVO₄:Eu (YVO) and Gd₂O₂S:Tb (Gadox), were chosen for their emission wavelengths and long decay times (around 500 µs), which are beneficial for counting mode operation.
  • Optical Cross-Talk (OCT) Exploitation: The paper demonstrates a technique to maintain detector linearity even at high radiation rates by strategically utilizing the SiPM's optical cross-talk. This is a novel approach to overcome a common limitation of SiPMs.
  • Laboratory Characterization: Extensive laboratory tests were performed using an X-ray beam to evaluate key dosimetric figures of merit, including Minimum Detectable Rate (MDR), linearity range, and sensitivity.
  • Pre-Clinical Trials: The system was tested with actual 192Ir (High Dose Rate - HDR) and 125I (Low Dose Rate - LDR) brachytherapy sources in water-equivalent phantoms. These trials aimed to assess compliance with the TG43-U1 dosimetry protocol, a standard in brachytherapy.
  • Data Analysis: Measurements were compared against theoretical predictions from the TG43-U1 protocol, with specific attention paid to decoupling angular dependence and addressing the "stem effect" (artifacts from radiation interacting with the fiber itself).

Datasets

Not explicitly discussed as "datasets" in the traditional machine learning sense. The paper relies on:

  • Experimental Data: Measurements obtained from laboratory irradiations using an X-ray cabinet and from pre-clinical trials with actual 192Ir and 125I brachytherapy sources.
  • Lookup Tables: Data from external sources, such as the physics.carleton.ca/clrp/seed_database, were used for anisotropy factors.
  • NIST Data: X-ray mass attenuation coefficients from NIST were used for calculations.

No new benchmarks were introduced; rather, the system's performance was evaluated against existing standards like the TG43-U1 protocol.

Key Insights and Ideas

  • OCT as a Feature, Not Just a Bug: The paper ingeniously proposes using optical cross-talk, often considered a drawback in SiPMs, as a mechanism to extend the linearity range of the detector. This is a significant conceptual shift.
  • SiPM Selection is Critical: The choice of SiPM significantly impacts performance, with different models excelling in sensitivity or linearity. The HAMAMATSU S13360-1375 showed higher sensitivity, while KETEK's PM1125 was favored for HDR due to its linearity.
  • Stem Effect is a Major Hurdle: The "stem effect," where radiation interacting with the fiber itself causes artifacts, is a significant challenge, particularly at close distances to the source. This effect leads to non-linearities and splits in the measured data.
  • TG43-U1 Compliance is Achievable (with caveats): The system demonstrates compliance with the TG43-U1 protocol, especially at distances greater than 2 cm for HDR. Overcoming the stem effect at closer distances is a key area for future work.
  • Real-time In-Vivo Monitoring is Feasible: The successful characterization and pre-clinical trials suggest that a real-time, in-vivo dosimetric system for brachytherapy is within reach.

Key Results

  • SiPM Performance: The HAMAMATSU S13360-1375 exhibited the highest sensitivity, while KETEK's PM1125 was identified as the optimal choice for HDR applications due to its linearity.
  • HDR Performance: The ORIGIN prototype for HDR met and exceeded design specifications, demonstrating a spatial resolution of 0.40 ± 0.03 mm at 5 cm, well within the project's 1 mm requirement. The MDR was found to be 2.95 ± 0.08% at 10 cm, exceeding the 5% specification.
  • LDR Performance: For LDR, the system's sensitivity was found to be limited by Dark Count Rate (DCR). However, the recorded PCR exceeded the required MDR for distances below 2.5 ± 0.1 cm.
  • TG43-U1 Compliance: The system showed TG43-U1 compliance at distances greater than 2 cm for HDR, with the stem effect being the primary cause of deviation at closer ranges. For LDR, the data supported TG43-U1 predictions.

Limitations and Future Work

  • Stem Effect Mitigation: The primary limitation identified is the "stem effect," which causes deviations from TG43-U1 compliance at close source-sensor distances, especially in HDR. Future work will focus on implementing solutions like fiber splitting and Cherenkov light filters to address this.
  • LDR Sensitivity: For LDR treatments, the sensitivity is limited by the DCR. The authors propose using thermo-electric cooled SiPMs and a custom optical system to reduce DCR by an order of magnitude, aiming for operation at -20 °C.
  • Angular Dependence in LDR: While not a major issue for LDR due to the source geometry, future work might explore more comprehensive angular dependence studies if needed.
  • System Integration: The paper focuses on the characterization of individual components and prototypes. Further work will involve integrating these into a complete, user-friendly system for clinical use.