TL;DR
This paper presents a novel, minimally invasive technique using a Foley catheter to create a temporary esophageal balloon occlusion during sedated gastroscopy in patients at high risk of aspiration. This method successfully prevented reflux and aspiration in a post-gastrectomy patient, offering a potentially safer alternative to intubation or awake procedures.
What Problem Does This Paper Address?
Upper endoscopy, while a common diagnostic procedure, carries a risk of pulmonary aspiration, particularly in patients with altered upper gastrointestinal anatomy (e.g., after gastrectomy) or severe reflux. These anatomical changes can compromise the lower esophageal sphincter, increasing the likelihood of gastric contents entering the airway during sedation. Current strategies to mitigate this risk include awake endoscopy (which is poorly tolerated), or general anesthesia with endotracheal intubation (which is resource-intensive and carries its own risks). There's a need for a simpler, less invasive method to protect the airway during deep sedation in these high-risk individuals.
Methods
The core of this paper is a novel application of a standard medical device: an 18-Fr silicone Foley catheter. The technique involves:
- Transnasal Insertion: The Foley catheter is inserted through the patient's nostril into the mid-esophagus. The depth is estimated using external measurements.
- Esophageal Balloon Occlusion: The catheter's inflatable cuff is filled with 20 ml of sterile saline, creating a temporary seal within the esophagus. This occludes the lumen, preventing reflux from the stomach.
- Continuous Suction: The catheter's drainage port is connected to wall suction at approximately -20 KPa. This actively aspirates any refluxate that might accumulate below the balloon, effectively decompressing the remnant stomach.
- Deep Sedation: The patient is then deeply sedated (MOAA/S score ≤2) using propofol and alfentanil, while spontaneous respiration is maintained.
- Controlled Deflation and Endoscopy: During the gastroscopy, the esophageal balloon is gradually deflated under direct endoscopic visualization. This allows for careful monitoring for any signs of reflux. If reflux is observed, the balloon can be reinflated to aspirate the contents before proceeding.
This approach is presented as an intermediate strategy between awake endoscopy and endotracheal intubation.
Datasets
Not discussed in the paper. This is a case report focusing on a single patient.
Key Insights and Ideas
- Creative Repurposing of Devices: The study highlights how a common and inexpensive device like a Foley catheter can be ingeniously adapted to solve a complex clinical problem, demonstrating the value of innovation in resource-limited scenarios.
- Dynamic Airway Protection: Unlike static methods, this technique allows for dynamic control of the esophageal barrier. The ability to monitor for reflux during balloon deflation and reinflate if necessary provides a more responsive approach to aspiration prevention.
- Bridging the Gap in Sedation Strategies: The technique offers a promising "middle ground" for high-risk patients who cannot tolerate awake procedures but for whom intubation might be considered excessive. It aims to balance patient comfort with robust airway protection.
- Potential for Minimally Invasive Airway Management: By avoiding endotracheal intubation, this method could reduce the risks associated with intubation, such as pharyngeal injury and prolonged recovery, while still enabling deep sedation.
Key Results
- The procedure was successfully performed on a 56-year-old male with a history of proximal gastrectomy and chronic reflux, who was deemed high-risk for aspiration.
- The esophageal balloon occlusion with continuous suction allowed for deep sedation without endotracheal intubation.
- The 8-minute gastroscopy was completed without any signs of aspiration, hypoxemia, coughing, or retching.
- The patient recovered uneventfully and was discharged the same day, reporting only mild nasal discomfort.
- Endoscopic visualization confirmed the presence of retained gastric contents in the stomach, which did not reflux into the esophagus during the procedure.
Limitations and Future Work
The authors acknowledge several limitations:
- Single-Case Report: The findings are based on a single patient, and causality cannot be definitively established. The absence of aspiration could be due to multiple factors.
- Potential for Mucosal Injury: While the occlusion time was short (4 minutes), prolonged inflation or overinflation could potentially injure the esophageal mucosa. Guidelines for tamponade devices recommend intermittent deflation and removal within 24 hours to minimize this risk.
- Discomfort and Technical Challenges: Transnasal insertion of a relatively large catheter can be uncomfortable. The technique requires close coordination between the anesthesiologist and endoscopist.
- Contraindications: The technique is not suitable for patients with esophageal lesions, foreign bodies, or active bleeding.
Future Work:
- Larger Studies: Further evaluation in larger studies is warranted to validate the safety and efficacy of this technique.
- Refining Parameters: Technical parameters such as optimal balloon size, inflation volume, and duration need to be refined.
- Defining Role: The role of this technique within the spectrum of airway management strategies for high-risk patients needs to be clearly defined.
- Advanced Confirmation: Future applications could incorporate bedside ultrasound and end-tidal carbon dioxide monitoring to confirm esophageal placement and rule out tracheal insertion.
Summary updated