Cardiac output monitors in septic shock: do they deliver what matters? A systematic review and meta-analysis

Raquel Lamarche-Fontaneto, Lavi Oud, Kevin Howell, Simran Ganeriwal, Gaurav Manek, Mary Schleicher, Jihad Mallat, Maurizio Cecconi, Abhijit Duggal, Matthew SiubaCritical 2025 · DOI

Abstract

This systematic review and meta-analysis evaluates the interchangeability of cardiac output (CO) monitoring devices against reference methods in adult patients with septic shock. Analyzing 26 studies (37 datasets, 1,323 patients), we found a pooled percentage error (PE) of 49%, far exceeding the 30% industry-standard threshold for agreement. While calibrated pulse contour analysis (PCA) showed acceptable agreement (PE 25%), uncalibrated and non-invasive methods performed poorly. Crucially, the study highlights a significant gap in clinical validation: current frameworks prioritize static accuracy over the real-time trending and precision that intensivists require for bedside decision-making.

Introduction

In the management of septic shock, maintaining adequate oxygen delivery is a dynamic challenge. Clinicians rely on cardiac output (CO) monitoring to guide fluid resuscitation and vasopressor titration. While the pulmonary artery catheter (PAC) remains a gold standard, its invasiveness has led to the proliferation of minimally invasive and non-invasive alternatives. However, the physiological chaos of sepsis—characterized by vasoplegia and rapid hemodynamic shifts—often undermines the algorithmic assumptions of these newer devices. This review investigates whether these monitors are truly interchangeable with reference methods and if they provide the "trending" data necessary for life-saving interventions.

Methodology

The authors conducted a comprehensive systematic search of electronic databases through January 2025. Prospective studies comparing CO monitors with PAC, transpulmonary thermodilution (TPTD), or echocardiography in septic shock patients were included. The primary metric for agreement was Percentage Error (PE), with a threshold of <30% defined as acceptable. Trending ability was assessed via concordance (target ≥90%). Data were pooled using the Sidik–Jonkman random-effects model to account for inter-study variability.

PRISMA flow diagram illustrating the systematic selection of studies for the meta-analysis.

Key Results and Data Synthesis

The meta-analysis encompassed 37 unique device-reference datasets. The pooled mean bias was low at 0.15 L/min, but the wide limits of agreement (±3.45 L/min) and a pooled PE of 49.4% indicate poor overall reliability.

Performance varied significantly by device category:

  • Calibrated PCA: The most reliable category, achieving a pooled PE of 24.7%.
  • Uncalibrated PCA: Demonstrated poor agreement, likely due to the inability of fixed algorithms to account for the vascular changes in sepsis.
  • Bioimpedance and Bioreactance: Showed the worst performance, with a pooled PE of 71.8%.
Summary of study characteristics and performance metrics for various cardiac output monitoring devices.

A critical finding of this review is the lack of data on trending ability. For an intensivist, the absolute CO value is often less critical than the direction and magnitude of change following a clinical intervention (e.g., a fluid bolus). Only 15 of the 37 datasets reported trending ability, and only three achieved the ≥90% concordance target.

Furthermore, the authors argue that the medical community is caught in a "Bland-Altman Trap." By focusing almost exclusively on static accuracy (how close a device is to a gold standard at a single point in time), validation studies often ignore time response (the lag in detecting changes) and precision (signal-to-noise ratio). In the fast-paced environment of the ICU, a monitor that is slightly less accurate but highly precise and responsive may be more clinically useful than a slow, accurate one.

Conclusion and Future Directions

Most current CO monitors fail to meet the standard criteria for interchangeability in septic shock. While calibrated PCA remains a viable tool, there is an urgent need to shift the validation paradigm. Future research must prioritize metrics that matter at the bedside: precision, time response, and trending ability. Until then, clinicians should interpret absolute CO values from minimally invasive monitors with caution, particularly in the highly unstable early phases of septic shock.

Summary updated

Abstract

To evaluate the interchangeability of cardiac output (CO) monitoring devices compared to reference methods in adult ICU patients with septic shock, we systematically searched electronic databases through January 2025 for prospective studies comparing CO monitors with pulmonary artery catheter (PAC), transpulmonary thermodilution (TPTD), or echocardiography. Eligible studies included Bland-Altman analysis and, when available, trending assessment via polar or 4-quadrant plots, precision, and time response. Agreement was defined as percentage error (PE) < 30%, and acceptable trending as concordance ≥ 90%. Pooled bias, limits of agreement (LoA), and PE were calculated using the Sidik-Jonkman random-effects model. Twenty-six studies were included, yielding 37 unique device-reference datasets and encompassing 1,323 patients. PAC was the most common reference (18 datasets), followed by TPTD (16) and echocardiography (3). The pooled bias was 0.15 L min⁻¹ with LoA of ± 3.45 L min⁻¹ and pooled PE of 49%. Calibrated pulse contour analysis (PCA) showed the best agreement (PE 25%), whereas uncalibrated PCA, thoracic electrical bioimpedance, and bioreactance demonstrated poor agreement (PE ≥ 52%). Heterogeneity for mean bias was high across all subgroups (I² >80%). Of 15 datasets reporting trending, only three achieved concordance ≥ 90%. Most CO monitors demonstrate poor agreement with reference methods in septic shock. However, their true clinical utility remains unclear, as usual validation frameworks-centered on Bland-Altman analysis-overlook metrics that matter most to intensivists. Precision, time response, and trending ability are critical for real-time decision-making but were rarely assessed. Future studies must incorporate these parameters to meaningfully evaluate device performance at the bedside. PROSPERO registration: CRD42024509384.