Calibration and monitoring of LHAASO-KM2A muon detectors with muon decay events

Xiong Zuo, Gang Xiao, Feng Shaohui, Xiurong Li, Cong Li, Ning Cheng, Lingyu Wang, Jinfan Chang, Wei Wang, M. H. Gu, Li Fei, Jia Liu, H. K. Lv, X. D. Sheng, H. H. He, Hongming Li, Guanghua GongProceedings of 35th International Cosmic Ray Conference — PoS(ICRC2017) 2017 · DOI

Abstract

The Large High Altitude Air Shower Observatory (LHAASO) will be constructed at 4400 m asl in Daocheng, Sichuan Province, with the aim of discovering sources of ultra-high energy cosmic rays and studying cosmic ray physics. The whole array includes 1171 muon detectors (MDs), constituting the largest muon detector array in the world. The special high altitude and wide-field environment requires simple and stable calibration as well as monitoring method of MDs. In this paper, the muon decay events will be selected for calibrating the charge and monitoring the water level of each MD. In the prototype muon detector, the simulation and experiment results reveal that the charge of the decay electron is approximately 0.147 times that of the vertical equivalent muon for a standard MD, and the $Q_{VEM}/Q_e$ shows a linear correlation with the water level.

Introduction

The Large High Altitude Air Shower Observatory (LHAASO), situated at 4400 m above sea level in Daocheng, Sichuan, is a major scientific endeavor aimed at uncovering the origins of ultra-high energy cosmic rays. A critical component of this facility is the KM2A array, which incorporates 1171 Muon Detectors (MDs). These detectors are essential for distinguishing between gamma-ray-induced and cosmic-ray-induced air showers. Given the remote, high-altitude location and the fact that these detectors are buried under 2.5 meters of soil for shielding, traditional manual calibration and monitoring are impractical. This paper presents a robust, automated method for the calibration and monitoring of MDs using muon decay events.

Detector System and Electronics

Each MD consists of a concrete tank containing a high-reflectivity liner and a water bag filled with ultra-pure water. A photomultiplier tube (PMT) is positioned at the top to detect Cherenkov light produced by passing particles. The signals are processed by an analog front end and digitized using Fast Analog-to-Digital Converters (FADCs) within an FPGA-based electronics system.

Schematic of the Muon Detector (MD) electronics and data acquisition system.

The primary challenges for long-term operation include maintaining the stability of the PMT gain and monitoring the water level within the tanks. Environmental factors and potential leaks in the water bags can compromise the detector's response, necessitating a reliable remote monitoring system.

Calibration Methodology: Muon Decay as a Standard Candle

The proposed method leverages the physics of low-energy muons that stop within the detector volume. These muons decay into two neutrinos and a Michel electron with a mean lifetime of approximately 2.2 microseconds. Because Michel electrons have a maximum energy of 53 MeV, they typically deposit their entire energy within the 1.2-meter water depth of a standard MD, regardless of the total water volume. This makes them an ideal "standard candle" for calibration.

To distinguish these faint decay signals from background noise, researchers utilize a timing variable, $T_c$, representing the mean hitting time of photons. This allows for the effective selection of "clustered" hits characteristic of decay electrons. The calibration relies on the ratio between the Vertical Equivalent Muon (VEM) charge ($Q_{VEM}$) and the Michel electron charge ($Q_e$). While $Q_{VEM}$ is directly proportional to the water level (track length), $Q_e$ remains stable, allowing environmental variables like water clarity or PMT aging to cancel out in the ratio.

Results and Performance

Simulation studies using GEANT4 and experimental data from prototype detectors at the Yangbajing Cosmic Ray Observatory have validated this approach. The results show a clear separation between the charge distributions of Michel electrons and VEM signals. For a standard MD with a 120 cm water level, the charge of a decay electron is approximately 0.147 times that of a VEM.

Charge distributions for muon decay events (left) and the linear correlation between the $Q_{VEM}/Q_e$ ratio and water level (right).

Crucially, the ratio $Q_{VEM}/Q_e$ exhibits a strong linear correlation with the water level $D$, described by the relationship:

$$\frac{Q_{VEM}}{Q_e} = (0.8307 \pm 0.0016)\% \cdot D(cm)$$

This linear dependence allows for the remote monitoring of water levels with an accuracy of approximately 1.3 cm. The method was successfully tested on a prototype detector with a known slow leak, where the $Q_{VEM}/Q_e$ ratio accurately tracked the decreasing water level over an eight-month period.

Conclusion

The use of muon decay events provides a simple, stable, and automated solution for the calibration and monitoring of the LHAASO-KM2A muon detector array. By utilizing Michel electrons as a reference, the system can ensure the long-term stability and accuracy of the world's largest muon detector array, even in the challenging environment of the Tibetan Plateau. This method is now being implemented across the full 1171-detector array to provide real-time health monitoring and data quality assurance.

Summary updated

Abstract

The Large High Altitude Air Shower Observatory (LHAASO) will be constructed at 4400 m asl in Daocheng, Sichuan Province, with the aim of discovering sources of ultra-high energy cosmic rays and studying cosmic ray physics. The whole array includes 1171 muon detectors (MDs), constituting the largest muon detector array in the world. The special high altitude and wide-field environment requires simple and stable calibration as well as monitoring method of MDs. In this paper, the muon decay events will be selected for calibrating the charge and monitoring the water level of each MD. In the prototype muon detector, the simulation and experiment results reveal that the charge of the decay electron is approximately 0.147 times that of the vertical equivalent muon for a standard MD, and the QVEM/Qe shows a linear correlation with the water level.
BibTeX
@inproceedings{zuo2017calibration,
  author = {Xiong Zuo and Gang Xiao and Feng Shaohui and Xiurong Li and Cong Li and Ning Cheng and Lingyu Wang and Jinfan Chang and Wei Wang and M. H. Gu and Li Fei and Jia Liu and H. K. Lv and X. D. Sheng and H. H. He and Hongming Li and Guanghua Gong},
  title = {{Calibration and monitoring of LHAASO-KM2A muon detectors with muon decay events}},
  booktitle = {Proceedings of 35th International Cosmic Ray Conference — PoS(ICRC2017)},
  year = {2017},
  doi = {10.22323/1.301.0441},
}
Figure