Axonal stimulation affects the linear summation of single-point perception in three Argus II users

Yuchen Hou, Devyani Nanduri, Jacob Granley, James D. Weiland, Michael BeyelerJournal of Neural Engineering 2024 · DOI

TL;DR

This study investigates how electrical stimulation in retinal implants creates perceived light spots (phosphenes) and how these combine. Researchers found that the shape of phosphenes from stimulating two electrodes simultaneously can be predicted by simply adding up the shapes from stimulating each electrode individually. This suggests a linear summation of perception and provides evidence for the "axon map model" of retinal stimulation.

What Problem Does This Paper Address?

Retinal implants aim to restore vision by electrically stimulating remaining retinal neurons to create phosphenes. While we know that the shape of a single phosphene depends on stimulation parameters (like amplitude and frequency) and where it's stimulated on the retina, how phosphenes combine when multiple electrodes are activated simultaneously is less understood. This is crucial for developing implants that can create more complex visual percepts, like patterns and objects, moving beyond simple dots of light. The paper specifically addresses whether phosphenes from paired stimulation can be predicted by the sum of individual phosphenes and how factors like electrode placement relative to nerve fiber bundles influence the perception of multiple phosphenes.

Methods

The researchers retrospectively analyzed phosphene drawings from three participants implanted with the Argus II Retinal Prosthesis. They used a dataset of 3548 drawings, where participants drew the perceived phosphenes on a touchscreen after electrical stimulation.

Key methods included:

  • Phosphene Shape Quantification: Phosphene shapes were analyzed using image processing techniques to extract descriptors like area, perimeter, major axis length, and minor axis length.
  • Linear Regression and Partial Correlation: These statistical methods were used to determine how stimulus parameters (amplitude, frequency) and neuroanatomical parameters (electrode-fovea distance, electrode-retina distance, between-axon distance, along-axon distance) predicted phosphene shape and the number of perceived phosphenes.
  • Axon Map Model Validation: The study specifically tested the "axon map model," which predicts that the number of perceived phosphenes is more influenced by the distance between nerve fiber bundles ("between-axon distance") than the distance along the axon itself ("along-axon distance").
  • Linear Summation Hypothesis Testing: The core hypothesis that paired-electrode phosphene shape is the sum of single-electrode phosphene shapes was tested using simple linear regression.

Datasets

The study utilized a dataset of 3548 phosphene drawings collected from three blind participants implanted with the Argus II Retinal Prosthesis. This dataset was collected as part of previous interventional feasibility trials. No new datasets were introduced.

Key Insights and Ideas

  • Linear Summation of Phosphene Shape: The most significant finding is that the shape of phosphenes elicited by stimulating two electrodes simultaneously can be accurately predicted by the sum of the shapes elicited by stimulating each electrode individually. This strongly supports the idea that the visual system processes these stimulations in a linear fashion, at least concerning shape.
  • Axon Map Model Support: The number of perceived phosphenes during paired stimulation was significantly influenced by the "between-axon distance" (the distance between nerve fiber bundles) but not the "along-axon distance." This provides strong empirical evidence for the axon map model, suggesting that the underlying neural architecture (nerve fiber bundles) is a critical factor in determining how many distinct percepts are generated.
  • Predictive Power of Neuroanatomy: Electrode-fovea distance (retinal eccentricity) was a significant predictor of phosphene shape across participants, suggesting that where stimulation occurs on the retina influences the perceived form.
  • Stimulus Amplitude's Role: Stimulus amplitude was also a key predictor of phosphene shape, generally leading to larger and more elongated phosphenes, which aligns with previous findings.

Key Results

  • Shape Prediction: Simple regression showed that each paired-electrode phosphene shape descriptor could be predicted by the sum of the corresponding single-electrode shape descriptors (p < .001).
  • Number of Phosphenes: The number of elicited phosphenes tended to increase with "between-axon distance" (p < .05) in two out of three participants, but not with "along-axon distance."
  • Predictors of Paired-Electrode Shape: Multiple regression revealed that paired-electrode phosphene shape was primarily predicted by stimulus amplitude and electrode-fovea distance (p < .05).

Limitations and Future Work

  • Limited Participants: The study involved only three participants, which limits the generalizability of the findings.
  • Paired Stimulation Only: The analysis focused on single and paired-electrode stimulation. Future work should investigate how phosphenes combine with stimulation of more than two electrodes to better understand complex pattern perception.
  • Linear Models: The current analysis relied on linear models. Future research could explore nonlinear machine learning models to capture more complex relationships.
  • Dynamic Stimulation: The study used static stimulation. Future work could investigate dynamic stimulation strategies for perceptual grouping.
  • Clinical Implications: The authors suggest that these findings can inform the optimal surgical placement of retinal implants to maximize between-axon distances and efficiently tile the axon map, rather than just the retinal surface. This could lead to improved visual outcomes for patients.

Summary updated

Abstract

<i>Objective.</i>Retinal implants use electrical stimulation to elicit perceived flashes of light ('phosphenes'). Single-electrode phosphene shape has been shown to vary systematically with stimulus parameters and the retinal location of the stimulating electrode, due to incidental activation of passing nerve fiber bundles. However, this knowledge has yet to be extended to paired-electrode stimulation.<i>Approach.</i>We retrospectively analyzed 3548 phosphene drawings made by three blind participants implanted with an Argus II Retinal Prosthesis. Phosphene shape (characterized by area, perimeter, major and minor axis length) and number of perceived phosphenes were averaged across trials and correlated with the corresponding single-electrode parameters. In addition, the number of phosphenes was correlated with stimulus amplitude and neuroanatomical parameters: electrode-retina and electrode-fovea distance as well as the electrode-electrode distance to ('between-axon') and along axon bundles ('along-axon'). Statistical analyses were conducted using linear regression and partial correlation analysis.<i>Main results.</i>Simple regression revealed that each paired-electrode shape descriptor could be predicted by the sum of the two corresponding single-electrode shape descriptors (<i>p</i> < .001). Multiple regression revealed that paired-electrode phosphene shape was primarily predicted by stimulus amplitude and electrode-fovea distance (<i>p</i> < .05). Interestingly, the number of elicited phosphenes tended to increase with between-axon distance (<i>p</i> < .05), but not with along-axon distance, in two out of three participants.<i>Significance.</i>The shape of phosphenes elicited by paired-electrode stimulation was well predicted by the shape of their corresponding single-electrode phosphenes, suggesting that two-point perception can be expressed as the linear summation of single-point perception. The impact of the between-axon distance on the perceived number of phosphenes provides further evidence in support of the axon map model for epiretinal stimulation. These findings contribute to the growing literature on phosphene perception and have important implications for the design of future retinal prostheses.
BibTeX
@article{hou2024axonal,
  author = {Yuchen Hou and Devyani Nanduri and Jacob Granley and James D. Weiland and Michael Beyeler},
  title = {{Axonal stimulation affects the linear summation of single-point perception in three Argus II users}},
  journal = {Journal of Neural Engineering},
  year = {2024},
  doi = {10.1088/1741-2552/ad31c4},
}
Figure