Automatic ultrasound image alignment for diagnosis of pediatric distal forearm fractures

Peng Liu, Yujia Hu, Jurek Schultz, Jinjing Xu, Christoph von Schrottenberg, Philipp Schwerk, Josephine Edith Pohl, Guido Fitze, Stefanie Speidel, Micha PfeifferInt. J. Comput. Assist. Radiol. Surg 2025 · DOI

TL;DR

This paper introduces an automated pipeline for aligning pediatric distal forearm ultrasound images, aiming to replace X-rays for fracture diagnosis. By accurately merging dorsal and palmar ultrasound views, the method enables precise fracture angulation measurement and classification, offering a radiation-free and child-friendly alternative.

What Problem Does This Paper Address?

Pediatric distal forearm fractures are common, and X-rays are the current standard for diagnosis. However, X-rays involve radiation exposure and can be uncomfortable for children, requiring specific positioning. Point-of-care ultrasound (POCUS) offers a radiation-free, more child-friendly alternative, but accurately assessing fracture angulation and displacement from POCUS images is challenging due to the need to combine views from different orientations. Manually aligning these views is time-consuming and impractical for routine clinical use. This paper addresses the need for an automated method to align POCUS images, making them a viable replacement for X-rays in diagnosing distal forearm fractures.

Methods

The core of the proposed solution is an automatic pipeline for aligning paired POCUS images of the same bone from opposite orientations (e.g., dorsal and palmar views of the radius).

  1. Bone Boundary Delineation: A deep learning model (nnUnet) is used to segment the bone boundaries.
  2. Landmark Extraction: Key anatomical landmarks are derived from the segmentation results. These include the physis (growth plate) and points along the diaphysis (bone shaft). Specifically, three feature points are identified for each image: the physis point ($L_p$), a point near the end of the diaphysis ($L_{d1}$), and a point further along the diaphysis ($L_{ao}$).
  3. Optimization-Based Alignment: These landmarks guide an optimization process to find a transformation matrix ($T$) that aligns one image ($I_m$) to a fixed reference image ($I_f$). Three constraints are employed:
  4. Latitudinal Constraint ($C_a$): Ensures that the Y-components of the physis points are within a specific range and their X-components are close.
  5. Longitudinal Orientation Constraint ($C_o$): Enforces that the orientation vectors of the diaphysis segments in both images are parallel.
  6. Longitudinal Position Constraint ($C_p$): Matches the bone width by ensuring the distance between the diaphysis line segments in both images, projected onto the normal vector of the fixed image, is consistent.
  7. Overall Optimization Target: A weighted sum of these three constraints forms the objective function, minimized using a Gradient Descent algorithm.

Datasets

  • Training Data: 40 images from 20 pediatric patients (healthy, buckle fracture, and greenstick fracture) were used to train the nnUnet segmentation model. Manual segmentation was performed using brush tools on the CVAT platform. The dataset was split into 36 images for training and 4 for testing.
  • Evaluation Data:
    • Alignment Experiment: 10 samples (If, Im, Ir - dorsal POCUS, palmar POCUS, and reference lateral X-ray) were used for manual alignment evaluation.
  • Morphology Experiment: 11 matched X-ray and POCUS images from the same patients were used for fracture classification and angulation measurement.

Key Insights and Ideas

  • POCUS as a Viable X-ray Alternative: The study strongly suggests that aligned POCUS images can provide sufficient information for diagnosing distal forearm fractures, potentially replacing X-rays.
  • Automated Alignment is Crucial: Manual alignment is tedious. The proposed automatic pipeline significantly streamlines the process, making POCUS a more practical diagnostic tool.
  • Multi-Constraint Optimization: Combining latitudinal, longitudinal orientation, and longitudinal position constraints is key to achieving accurate and robust alignment, preventing local minima.
  • Landmarks from Segmentation: Leveraging deep learning for segmentation to extract robust anatomical landmarks is an effective strategy for guiding the alignment process.

Key Results

  • Alignment Accuracy: The alignment experiment showed a mean registration error of 2.66 mm (Hausdorff distance) and 1.05 mm (Chamfer distance) compared to reference lateral X-rays.
  • Fracture Classification: In the morphology experiment, blinded evaluators achieved comparable accuracy on merged POCUS images (mean accuracy of 0.76) to X-ray images (mean accuracy of 0.97), with two out of three participants achieving 0.82 accuracy with POCUS.
  • Angulation Measurement: Mean absolute errors (MAEs) for angulation measurement with merged POCUS were 4.4°, 3.7°, and 5.0° for the three participants, which are larger than X-ray MAEs (2.8°, 1.8°, and 2.0°), but the discrepancy is considered practical.
  • Surgeon Feedback: Initial feedback from surgeons indicated that the aligned POCUS results were often sufficient for diagnosis.

Limitations and Future Work

  • Landmark Dependency: The pipeline's performance is highly dependent on the accuracy of detected landmarks, which can be challenging in cases with high noise or extreme fracture presentations.
  • Image Plane Misalignment: POCUS image pairs may not always be captured in perfectly corresponding planes, and deviations in collection orientation can lead to ambiguity in alignment.
  • X-ray Projection Issues: Reference X-rays themselves can have projection issues affecting physical scale accuracy.
  • Small Cohort Size: The study was conducted on a small cohort (11 samples for morphology experiment), which limits the statistical significance of some findings (e.g., correlation analysis).
  • Future Work: The authors plan to:
  • Increase the dataset size to improve diagnostic accuracy and reliability.
  • Focus on collecting more POCUS images.
  • Improve performance in detection and alignment.
  • Enhance robustness against ultrasound noise.

Summary updated