Automatic system for high-throughput and high-sensitivity diagnosis of SARS-CoV-2

Jun Lu, Weihua Fan, Zihui Huang, Ke Fan, Jianhua Dong, Jisheng Qin, Jianzhong Luo, Zhizhong Zhang, Guodong Sun, Chaohui Duan, Kunyi Pan, Wenshen Gu, Xiao ZhangBioprocess 2022 · DOI

Introduction

The COVID-19 pandemic highlighted a critical need for rapid, high-throughput, and sensitive diagnostic tools. While Reverse Transcription Polymerase Chain Reaction (RT-PCR) remains the gold standard, its reliance on complex thermal cycling and skilled labor creates significant bottlenecks. Researchers from the Guangzhou Institutes of Biomedicine and Health have developed a fully automated robotic system that streamlines the entire diagnostic pipeline—from sample lysis to data reporting—using isothermal Cross-Priming Amplification (CPA).

Addressing the Testing Power Gap

Mass population screening requires a balance between speed, sensitivity, and safety. Manual RT-PCR is slow and exposes lab personnel to infection risks. Conversely, many point-of-care tests sacrifice sensitivity for speed. This robotic system addresses this "testing power" gap by providing a "specimen-in, data-out" solution that minimizes human intervention while maintaining high sensitivity for low-load carriers.

The automated diagnostic workflow from patient sample to result reporting.

System Design and Methodology

The system is a 25-axis modular robotic assembly line controlled by a custom ARM+DSP+FPGA integrated multi-core board. Key innovations include:

  • Robotic Precision: Linear motors achieve <50 µm accuracy. A 5th-order polynomial "S-curve" acceleration algorithm ensures smooth movement, preventing liquid splashing and mechanical wear.
  • Isothermal Amplification (CPA): By operating at a constant 58 °C, the system avoids the hardware complexity of thermal cycling, speeding up the reaction.
  • Magnetic Rod Purification: A 96-format magnetic rod matrix extracts nucleic acids from 200 µl specimens.
  • Enhanced Detection: Photomultiplier Tube (PMT) detectors provide single-photon counting, significantly improving the signal-to-noise ratio compared to standard CMOS cameras.
  • CRISPR Integration: A two-step CRISPR-based assay can be triggered for ultra-high sensitivity, detecting as few as 50 viral copies.
Internal deck layout and hardware components of the robotic system.

Results and Performance

The system was validated using 1,129 clinical specimens and compared against CDC-approved RT-PCR kits.

  • Throughput: The robot processes 192 samples per batch in under 180 minutes, enabling a daily capacity of 800–1,000 samples with minimal staffing.
  • Sensitivity: One-step CPA achieved a limit of detection (LoD) of 100 copies/reaction, which improved to 50 copies/reaction with the CRISPR-based probe.
  • Accuracy: Clinical trials showed substantial agreement (KAPPA test: κ=0.73) with RT-PCR. Discrepancies were primarily limited to samples with viral loads near the LoD.
  • Automation: Weighted Mean Gradients (WMGs) are used to automatically distinguish positive from negative samples, eliminating human bias.

Discussion and Future Directions

The system's "specimen-in" design, combined with integrated HEPA filtration and UV sterilization, significantly enhances biosafety. While currently optimized for large batches, future iterations aim to improve flexibility for smaller sample sizes. Additionally, the authors plan to implement multi-gene detection (up to four genes) to differentiate between SARS-CoV-2 and other respiratory viruses like Influenza. The CRISPR-based module also offers a pathway for rapid variant screening, such as identifying Delta or other variants of concern.