Automatic Planning Tools for Lumbar Pedicle Screws: Comparison and Validation of Planning Accuracy for Self-Derived Deep-Learning-Based and Commercial Atlas-Based Approaches

Moritz Scherer, Lisa Kausch, Akbar Bajwa, Jan‐Oliver Neumann, Basem Ishak, Paul Naser, Philipp Kickingereder, Karl Kiening, Klaus Maier‐Hein, Andreas UnterbergJournal of Clinical Medicine 2023

TL;DR

This study compares a novel deep learning (DL) based tool for automated lumbar pedicle screw planning against a commercial atlas-based (ATL) system. The DL approach demonstrated superior accuracy and robustness, particularly in cases with complex spinal anatomy, suggesting its potential to revolutionize surgical planning.

What Problem Does This Paper Address?

Planning the placement of pedicle screws in the lumbar spine is a critical step in spinal surgery, directly impacting patient safety and surgical outcomes. Traditionally, this planning is done manually, which is time-consuming and requires significant surgical expertise. While automated planning tools aim to expedite this process, their accuracy can be compromised by variations in patient anatomy, such as degenerative changes, deformities, or prior surgeries. This study addresses the need for automated planning tools that are not only accurate but also robust to these anatomical complexities, which are common in real-world surgical scenarios. The authors aim to compare the performance of a self-developed deep learning tool against a commercially available atlas-based system, evaluating their accuracy and reliability across different levels of spinal anatomical alteration.

Methods

The researchers conducted an ex vivo study using CT scans from 50 patients who underwent lumbar and sacral instrumentation. They compared three planning approaches:

  1. Manual Planning: Performed by two experienced spine surgeons to establish a "ground truth" reference.
  2. Deep Learning (DL) Based Planning: A novel, self-derived tool integrated into the Medical Imaging and Interaction Toolkit (MITK). This tool uses a deep neural network (nnU-Net) trained on a large dataset of virtually planned screws to predict screw trajectories and dimensions directly from CT images. It operates as an image segmentation task, identifying desired screw locations.
  3. Commercial Atlas-Based (ATL) Planning: Utilized Brainlab's Elements Spine & Trauma 3D screw planning app. This approach relies on registering patient-specific image data to an anatomical atlas to derive screw plans.

The study evaluated planning accuracy in 3D space by calculating the minimal absolute distance (MAD) between the planned screw head and tip points and the reference screws, as well as the angular deviation of the screw direction. They also performed a qualitative evaluation using the Gertzbein-Robbins (GR) classification to assess screw placement relative to cortical bone margins. A key aspect of their analysis was comparing the performance of the automated tools not just against the manual reference but also in the context of interrater variability of manual planning, providing a more realistic benchmark for clinical utility. Furthermore, they stratified cases into three categories based on the degree of spinal anatomy alteration and alignment to assess the robustness of each planning tool.

Datasets

The study utilized a dataset of anonymized CT scans from 50 patients who underwent lumbar and sacral instrumentation between January 2010 and December 2018. This dataset comprised a total of 256 screws across L1-L5 levels. The DL tool was previously trained on a dataset of 155 spine CTs with 1052 manually labeled screw trajectories. The study did not introduce a new benchmark dataset but rather used existing clinical data for evaluation.

Key Insights and Ideas

  • Deep Learning's Robustness: The most significant insight is the remarkable robustness of the DL approach to anatomical variations and spinal alignment issues. This suggests DL can handle the complexities of real-world surgical cases far better than traditional atlas-based methods.
  • ATL's Susceptibility to Complexity: The ATL system struggled significantly with altered spinal anatomy, leading to a high rate of planning failure and a greater need for manual revision. This highlights a fundamental limitation of atlas-based methods when faced with deviations from idealized anatomical models.
  • Clinical Relevance of Interrater Variance: By comparing automated tool accuracy to interrater variance in manual planning, the study provides a more clinically relevant benchmark. It shows that the DL tool's performance is comparable to the variability seen between expert surgeons, indicating its potential to replace or augment manual planning.
  • L5 as a Challenge: Both DL and ATL planning tools encountered challenges at the L5 vertebra, suggesting this segment presents unique difficulties for automated screw placement, possibly due to its anatomical characteristics or common degenerative changes.

Key Results

  • Planning Success Rates: The DL tool achieved 100% planning success (256/256 screws), while the ATL tool succeeded in 81% of cases (208/256 screws), with 19% of screws failing to be planned.
  • Accuracy Metrics:
    • Screw Direction Deviation: DL (4.46 ± 2.86°) was comparable to interrater variance (5.27 ± 3.20°, p=0.07), while ATL (6.70 ± 3.53°) was significantly worse (p<0.001).
  • Screw Head Point Deviation: DL (3.93 ± 2.08 mm) was significantly better than ATL (7.77 ± 3.65 mm) compared to manual planning (interrater variance: 4.89 ± 2.04 mm, p<0.001 for DL vs. ATL).
  • Screw Tip Point Deviation: DL (3.49 ± 1.80 mm) was significantly better than ATL (7.81 ± 4.75 mm) compared to manual planning (interrater variance: 4.36 ± 2.25 mm, p=0.007 for DL vs. manual, p<0.001 for ATL vs. manual).
  • Qualitative Evaluation (GR Grade A/B): DL planning resulted in 97% GR-A and 3% GR-B screws. ATL planning had a higher proportion of non-GR-A screws (35% GR-B to E) and a significant failure rate (19%).
  • Revision Rates: 6% of DL-planned screws required revision (primarily due to proximal facet violations), whereas 55% of ATL-planned screws required revision due to GR violations, anterior breaches, or planning failures.
  • Robustness to Anatomy: In cases with severely altered spinal anatomy (Category III), ATL planning failure was pronounced (34%), while DL planning accuracy remained comparable to manual planning across all anatomical categories.

Limitations and Future Work

The authors acknowledge several limitations:

  • Ex Vivo Design: The study was conducted ex vivo, which may not perfectly replicate the dynamic environment of live surgery.
  • Small Dataset Size: While 50 cases and 256 screws are substantial, a larger dataset could further enhance generalizability.
  • Interrater Variability: While they used interrater variance as a benchmark, the manual planning was performed by highly experienced surgeons, potentially leading to lower interrater variability than in a broader surgical population.
  • Heterogeneity of Pathologies: The stratification was broad, and further evaluation in matched cohorts for specific conditions (e.g., increasing spondylolisthesis) is needed.

Future work suggested includes:

  • Explainability of DL: Developing methods to understand the DL algorithm's decision-making process to build trust and facilitate surgical decision-making.
  • Segment-Specific DL Models: Training DL models specifically for challenging segments like L5 to improve accuracy.
  • Incorporating Bone Mineral Density: Enhancing the DL framework to consider local bone mineral density for optimizing screw dimensions and maximizing fastening strength.
  • Clinical Trials: Moving towards prospective clinical trials to validate these findings in real-time surgical settings.

Summary updated

Abstract

Deep learning appears to be a promising approach to reliable automated screw planning, coping well with anatomic variations of the spine that severely limit the accuracy of ATL systems.
BibTeX
@article{scherer2023automatic,
  author = {Moritz Scherer and Lisa Kausch and Akbar Bajwa and Jan‐Oliver Neumann and Basem Ishak and Paul Naser and Philipp Kickingereder and Karl Kiening and Klaus Maier‐Hein and Andreas Unterberg},
  title = {{Automatic Planning Tools for Lumbar Pedicle Screws: Comparison and Validation of Planning Accuracy for Self-Derived Deep-Learning-Based and Commercial Atlas-Based Approaches}},
  journal = {Journal of Clinical Medicine},
  year = {2023},
  doi = {10.3390/jcm12072646},
}