Automated placement of stereotactic injections using a laser scan of the skull

Margaret M. Henderson, Vadim Pinskiy, Alexander Tolpygo, Stephen Savoia, Pascal Grange, Partha P. MitrabioRxiv 2014

TL;DR

This paper presents a novel software tool that uses 3D laser scans of mouse skulls to automate stereotactic injection targeting. By registering individual skull scans to a reference, the system significantly improves injection accuracy compared to traditional methods relying solely on skull landmarks, paving the way for more precise neuroscience experiments.

What Problem Does This Paper Address?

Stereotactic surgery is a cornerstone technique in experimental neuroscience, enabling researchers to precisely target specific brain regions for interventions like injections of tracers or drugs. The standard method for targeting these injections in mice relies on identifying anatomical landmarks on the skull, primarily the bregma and lambda. However, this approach suffers from inherent limitations:

  • Skull Variability: Individual mice exhibit variations in skull shape and curvature, leading to an imperfect correspondence between skull landmarks and actual brain locations.
  • Operator Error: Manual identification of these landmarks can introduce bias and inaccuracies.
  • Brain-Skull Mismatch: The reference atlases used for targeting are often based on post-mortem, processed brains, which can undergo deformation and shrinkage, further complicating the translation of skull coordinates to in-vivo brain coordinates.

These limitations collectively reduce the precision of stereotactic injections, potentially leading to off-target delivery, requiring repeat injections, and increasing experimental costs and time. The authors aim to overcome these challenges by developing a more robust and automated targeting system.

Methods

The core of this work is a software tool developed in MATLAB using the GUIDE environment. The methodology involves the following key steps:

  1. 3D Skull Scanning: A 3D laser scan of an individual mouse skull is acquired in vitro using a Keyence CCD laser displacement sensor. The resolution of these scans can range from 25x25 to 250x250 points.
  2. Initial Alignment: The scanned skull data is initially filtered for outliers and roughly aligned to a pre-existing reference skull. This initial alignment is guided by the manually identified bregma point.
  3. Symmetry Plane Determination: The symmetry plane of the scanned skull data is determined, which is then used to calculate initial rotations (theta and phi) around the y and z axes.
  4. Sagittal Crest Alignment: The sagittal crest line of the scanned skull is aligned with that of the reference skull. This alignment uses "eikonal distance" as an error metric, which measures the shortest distance from a point to a surface. This step defines the x-axis rotation (nu) and translations in the y and z directions.
  5. Point Cloud Dilation: Finally, dilation in the x, y, and z directions is optimized by minimizing the average eikonal distance across the entire point cloud.
  6. Transformation Application: Once the sample skull is registered to the reference skull, the parameters of this transformation are used to calculate a corrected injection site. This involves applying the inverse of the transformation to target coordinates derived from a brain atlas (like the Allen Brain Atlas).

The novelty lies in the automated registration of a 3D skull scan to a reference skull using a point cloud matching algorithm and eikonal distance as a metric, thereby creating a more accurate coordinate system for targeting.

Datasets

The study utilized CT scans of mouse skulls. Specifically, they obtained CT scan data with a 25 µm voxel size from the Henkelman lab at the Hospital for Sick Children in Toronto. The skull scan data used for testing the GUI were from male mice of strain C57BL/6J, aged 53±4 days and weighing 18.8-26.4 grams. The paper does not introduce new benchmark datasets but rather uses existing skull scan data for evaluation.

Key Insights and Ideas

  • Skull Geometry as a Proxy for Brain Location: The fundamental idea is that the shape and curvature of the skull are strong predictors of the underlying brain structures. This is supported by studies showing a tight fit and coordinated development between the mouse brain and skull.
  • Moving Beyond Fixed Landmarks: The system's ability to register individual skulls to a reference allows it to compensate for individual skull variations, a significant improvement over relying solely on fixed points like bregma.
  • Eikonal Distance for Surface Matching: The use of eikonal distance as an error metric for aligning the sagittal crest line is a clever way to quantify the similarity between skull surfaces.
  • Potential for Automation: The development of this software tool represents a significant step towards automating the entire process of stereotactic targeting, reducing manual intervention and potential errors.

Key Results

  • Registration Accuracy: The automated software was capable of registering sample skulls with an average eikonal distance error of 89.09 microns after fitting. This represents a statistically significant decrease from the pre-fitting average error of 265.76 microns.
  • Injection Targeting Accuracy: In a test on a single mouse, the automated protocol was able to target injections with an error of approximately 500 microns. Specifically, actual injection sites were offset from the Allen Brain Atlas (ABA) coordinates by 0.53 mm in the left hemisphere and 0.28 mm in the right hemisphere.
  • Functional Software: The software successfully ran on multiple data sets and performed registrations, demonstrating its functional capability.

Limitations and Future Work

  • In Vitro Scanning: The current method requires scanning the skull in vitro (after excision). A significant future step is to develop a protocol for in vivo scanning.
  • Brain Atlas Accuracy: The paper acknowledges that current brain atlases (like ABA) are based on post-mortem brains, which can be deformed. Future work includes developing a brain atlas that more accurately represents the in vivo mouse brain shape and position, potentially by registering MRI scans of extracted brains to the ABA.
  • Registration Accuracy Threshold: Determining the precise level of registration accuracy needed for effective injection planning is an important future step.
  • Improving the Algorithm: Further refinement of the automated registration algorithm is planned to achieve higher levels of accuracy.
  • Adaptability to Other Animals: The authors suggest that with an appropriate atlas, this protocol could be adapted for stereotactic targeting in other rodents like rats.

Summary updated

Abstract

Abstract Stereotactic targeting is a commonly used technique for performing injections in the brains of mice and other animals. The most common method for targeting stereoscopic injections uses the skull indentations bregma and lambda as reference points and is limited in its precision by factors such as skull curvature and individual variation, as well as an incomplete correspondence between skull landmarks and brain locations. In this software tool, a 3D laser scan of the mouse skull is taken in vitro and registered onto a reference skull using a point cloud matching algorithm, and the parameters of the transformation are used to position a glass pipette to place tracer injections. The software was capable of registering sample skulls with less than 100 micron error, and was able to target an injection in a mouse with error of roughly 500 microns. These results indicate that using skull scan registration has the potential to be widely applicable in automating stereotactic targeting of tracer injections.
BibTeX
@misc{henderson2014automated,
  author = {Margaret M. Henderson and Vadim Pinskiy and Alexander Tolpygo and Stephen Savoia and Pascal Grange and Partha P. Mitra},
  title = {{Automated placement of stereotactic injections using a laser scan of the skull}},
  year = {2014},
  doi = {10.1101/010603},
  howpublished = {bioRxiv},
}
Figure