Automated detection of neonatal pulmonary hypertension in echocardiograms with a deep learning model

Holger Michel, Ece Özkan, Kieran Chin-Cheong, Anna Badura, Verena Lehnerer, Stephan Gerling, Julia E. Vogt, Sven WellmannPediatric 2025

Abstract

This study presents a deep learning model that enables accurate, automated detection of pulmonary hypertension (PH) in infants using standard echocardiography videos, enhanced and evaluated with the systolic eccentricity index (EIs), an established and prognostically relevant echocardiographic parameter. The parasternal short-axis view showed the best single-view performance, while combined views further improved diagnostic accuracy. Explainability through saliency maps supports clinical acceptance by highlighting physiologically relevant regions in the decision process. This research adds novel evidence to the literature, demonstrating the utility of spatio-temporal convolutional neural networks for early, non-invasive diagnosis. The model provides a scalable and reproducible tool for routine PH screening, potentially improving early detection and clinical outcomes.

Introduction

Pulmonary hypertension (PH) in neonates and infants is a high-stakes condition linked to significant morbidity and mortality, often arising from respiratory failure or developmental lung diseases. While echocardiography is the primary non-invasive diagnostic tool, its interpretation is notoriously operator-dependent. It requires specialized expertise to capture the correct views and manually calculate indices like the systolic eccentricity index (EIs)—a measure of how much high pressure in the right ventricle flattens the interventricular septum (the wall between the heart's lower chambers).

There is a critical need for automated, objective, and scalable screening tools that can assist clinicians, especially in settings where pediatric cardiology expertise may not be immediately available around the clock. This study evaluates a deep learning framework designed to automate this diagnostic process using standard clinical echocardiography loops.

Methodology

Model Architecture

The researchers developed a pipeline to process standard 2D echocardiography videos, comparing two distinct technical approaches:

  • Spatial Approach (2D-CNN): A model focusing on individual frames to identify structural heart geometry changes.
  • Spatio-Temporal Approach (3D-CNN): A ResNet-18-based model that integrates motion information across multiple frames to capture the dynamics of the cardiac cycle.

Diagnostic Strategy

The model was evaluated across five standard echocardiographic views: Apical four-chamber (A4C), Parasternal long-axis (PLAX), and three Parasternal short-axis views (PSAX-P, PSAX-S, PSAX-A). A hierarchical majority voting strategy was implemented: first aggregating predictions across video sequences for a "view-level" prediction, and then combining multiple views for a final patient-level diagnosis.

Data and Labeling

The study utilized a retrospective dataset from the University of Regensburg, divided into a training/validation set (199 echocardiograms, 975 videos) and a held-out test set (77 echocardiograms, 378 videos). Ground truth labels were derived from quantitative EIs measurements: patients were classified as PH if EIs < 0.82 and healthy if EIs ≥ 0.87.

To ensure clinical interpretability, the team employed Grad-CAM (Gradient-weighted Class Activation Mapping) to generate saliency maps. These heatmaps highlight the anatomical structures most influential to the model’s decision, allowing for expert verification of the AI's focus.

Results and Discussion

Performance Metrics

The model achieved high diagnostic accuracy, particularly when integrating multiple views:

  • Single-View Performance: The PSAX-P view (at the papillary muscle level) emerged as the most effective single view, achieving an AUROC of 0.94 in validation and 0.93 in the held-out set. This finding is significant as it mirrors the clinical standard for manual EIs measurement.
  • Multi-View Integration: Combining three standard views (MV-3) further improved performance, reaching a validation AUROC of 0.96.
  • Robustness: On the independent held-out test set, the multi-view spatial approach maintained a strong AUROC of 0.90, with a balanced accuracy of 84%.

Insights into Cardiac Imaging

A notable finding was that 2D spatial models performed comparably to 3D spatio-temporal models on the test set. This suggests that the structural deformation of the heart—specifically the flattening of the septum—is a more dominant diagnostic signal than temporal motion patterns. Furthermore, saliency maps consistently highlighted the interventricular septum and left atrial filling areas, confirming that the model focuses on physiologically relevant features rather than image artifacts.

Conclusion and Future Work

This deep learning framework offers a promising tool for the automated detection of neonatal PH, providing objective and reproducible results that align with clinical intuition. While the model demonstrates high accuracy, future work should focus on:

  • Addressing the "Gray Zone": Evaluating performance on borderline cases (EIs 0.82–0.87) currently excluded from the study to ensure full clinical utility.
  • Multi-Center Validation: Testing the model across different hospitals and ultrasound manufacturers to ensure generalizability.
  • Severity Assessment: Developing capabilities to quantify PH severity rather than providing a binary classification.
  • Multimodal Data: Integrating additional metrics, such as Doppler flow velocities, to further refine diagnostic precision.

Summary updated

BibTeX
@inproceedings{michel2025automated,
  author = {Holger Michel and Ece Özkan and Kieran Chin-Cheong and Anna Badura and Verena Lehnerer and Stephan Gerling and Julia E. Vogt and Sven Wellmann},
  title = {{Automated detection of neonatal pulmonary hypertension in echocardiograms with a deep learning model}},
  booktitle = {Pediatric},
  year = {2025},
  doi = {10.1038/s41390-025-04404-3},
}
Figure