Augmented Reality Guidance for Needle Biopsies: A Randomized, Controlled Trial in Phantoms
Lists
Only you see this
Abstract
We report the results of a randomized, controlled trial to compare the accuracy of standard ultrasound-guided needle biopsy to biopsies performed using a 3D Augmented Reality (AR) guidance system. Fifty core biopsies of breast phantoms were conducted by a board-certified radiologist, with each set of five biopsies randomly assigned to one of the methods. The raw ultrasound data from each biopsy was recorded. Another board-certified radiologist, blinded to the actual biopsy guidance mechanism, evaluated the ultrasound recordings and determined the distance of the biopsy from the ideal position. A repeated measures analysis of variance indicated that the head-mounted display method led to a statistically significantly smaller mean deviation from the desired target than did the CRT display method. (2.48mm for control versus 1.62mm for augmented reality, p < 0.02). This result suggests that AR systems can offer improved accuracy over traditional biopsy guidance methods.
BibTeX
@inproceedings{rosenthal2001augmented,
author = {Michael Rosenthal and Andrei State and Joohi Lee and Gentaro Hirota and Jeremy Ackerman and Kurtis Keller and Etta D. Pisano and Michael R. Jiroutek and Keith E. Muller and Henry Fuchs},
title = {{Augmented Reality Guidance for Needle Biopsies: A Randomized, Controlled Trial in Phantoms}},
booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
year = {2001},
url = {https://openreview.net/forum?id=KbksCrm2ho},
}Loading related content...