Audio monitoring of bone cement disintegration in pulsating fluid jet surgery under laboratory conditions

Melanie Schaller, Sergej Hloch, Akash Nag, Dagmar Klichová, Frank Pude, Michal Zeleňák, Nick Janssen, Bodo RosenhahnResults 2025 · DOI

TL;DR

This paper introduces a novel approach to bone cement removal during orthopedic revision surgeries using a pulsating fluid jet combined with advanced audio monitoring. By leveraging State Space Models (SSMs) and analyzing the sound generated during the process, the system can accurately predict the erosion depth, overcoming visibility issues caused by splashing and paving the way for more precise and minimally invasive procedures.

What Problem Does This Paper Address?

Removing bone cement during revision surgeries is a critical but challenging task. Traditional manual methods can be imprecise, time-consuming, and potentially damaging to surrounding bone tissue, which is crucial for future implant fixation. Existing fluid jet techniques, while promising, often suffer from limited visibility due to splashing, making real-time monitoring and parameter adjustment difficult. This lack of precise control hinders the optimization of minimally invasive procedures, especially when using long nozzles required for such approaches. The paper aims to bridge this gap by developing a system that can accurately monitor and predict bone cement disintegration in real-time, even with limited visual feedback.

Methods

The core of this research lies in the integration of a pulsating fluid jet device with an advanced audio monitoring system powered by State Space Models (SSMs), specifically the S4D-Bio model.

  • Pulsating Fluid Jet: This technology uses a high-pressure fluid jet that is ultrasonically vibrated to create pulsating clusters of fluid. This method is designed to be cold and precise, minimizing damage to surrounding tissues. The study specifically investigates the efficacy of a novel, in-house designed "ultra-long nozzle" to enable more minimally invasive procedures.
  • Audio Monitoring: Instead of relying on visual feedback, which is obscured by splashing, the system captures the acoustic signals generated by the fluid jet interacting with the bone cement. A high-resolution microphone is used for this purpose.
  • State Space Models (SSMs) - S4D-Bio: The captured audio data is processed into Mel Spectrograms and fed into the S4D-Bio model. SSMs are adept at modeling dynamic systems and temporal dependencies. The S4D-Bio model, an adaptation of the S4 model, is designed to efficiently process these sequential audio features and predict the corresponding erosion profiles of the bone cement. This is the first known application of SSMs in this specific surgical technology context.
  • Data Correlation: The system correlates the audio signals with the physical erosion profiles of the bone cement, allowing for real-time prediction of the disintegration depth.

Datasets

The study generated a comprehensive dataset specifically for this research.

  • Bone Cement Samples: Samples were prepared to mimic clinical conditions, with intentionally uneven surfaces to simulate realistic challenges.
  • Erosion Profiles: Disintegration depth was measured using an optical microscope and processed into digital x, y, and z-axis data. Multiple profiles were taken for each experimental condition to ensure statistical reliability.
  • Audio Data: High-resolution audio recordings were captured synchronously with the erosion process.
  • Combined Dataset: The dataset consists of numerical ablation profiles and corresponding audio files. The paper states this dataset will be made publicly available under IEEE Dataport, along with the necessary scripts for loading and preprocessing.

Key Insights and Ideas

  • Audio as a Reliable Indicator: The research demonstrates that sound is a surprisingly effective proxy for monitoring bone cement disintegration, even in the presence of splashing that obscures visual cues. This opens up new avenues for non-invasive surgical monitoring.
  • SSMs for Surgical Prediction: The successful application of SSMs (S4D-Bio) in this context is a significant novelty. It highlights the potential of advanced machine learning models to handle complex spatio-temporal data in real-world surgical scenarios.
  • Long Nozzles for Minimally Invasive Surgery: The study validates the effectiveness of the novel long nozzle design, confirming its role in enabling more minimally invasive bone cement removal.
  • Bell-Shaped Erosion Curve: The erosion depth exhibits a bell-shaped curve with respect to the standoff distance, indicating an optimal distance for maximum efficiency and highlighting the importance of precise parameter control.
  • Noise Tolerance: The S4D-Bio model demonstrates a notable robustness to noise, which is a common challenge in real-world surgical environments.

Key Results

  • High Prediction Accuracy: The S4D-Bio model achieved a validation accuracy of 98.93% in predicting erosion profiles from audio data. This accuracy is defined as the percentage of correctly predicted points within an error threshold of 1 µm.
  • Effective Bone Cement Removal: The pulsating fluid jet technology, particularly with the long nozzle, proved effective for bone cement removal, offering a cold and precise alternative to manual methods.
  • Optimal Standoff Distance: The study identified an optimal standoff distance of approximately 5 mm, yielding a maximal disintegration depth of around 1327.71 ± 405.06 µm.
  • Benchmarking Performance: The S4D-Bio model significantly outperformed GRU and shallow MLP models in accuracy, with Deep-MLP showing comparable but slightly lower performance.

Limitations and Future Work

  • Laboratory Conditions: The current experiments were conducted under controlled laboratory conditions. Future work needs to involve in vivo validation to assess the system's performance in actual surgical environments.
  • Nozzle and Environment Optimization: Further research is needed to optimize nozzle configurations for different bone cement types and to adapt the acoustic monitoring technique for varying surgical environments.
  • Tissue Interaction: Investigations into potential tissue interactions at different frequencies are warranted to ensure the safety of the technology across diverse procedures.
  • Broader Applications: The methodological framework developed, particularly the noise-handling strategies and the SSM approach, has broader implications for other engineering disciplines and noisy environments. The study suggests potential for adaptation to other surgical specializations requiring non-invasive precision.

Summary updated

Abstract

• Demonstrated the time-reducing and bone damage-reducing effects of pulsating fluid jet technology when utilized for bone cement removal. • Demonstrated the efficacy of an in-house designed long nozzle, that is designed for minimally invasive surgical procedures. • Introduced the first application of SSMs in pulsating fluidjet surgery technology. • Introduced a novel audio monitoring technique for pulsating fluidjets to compensate for splashing effects during the erosion process of bone cement removal. • First to concatenate erosion profiles with audio data, introducing a novel metric to enhance the prediction of erosion profiles using audio data. • Provided an open-source dataset on erosion depth and audio signals to enhance surgical planning. • Demonstrated the efficacy of state space models for erosion profile prediction. This study investigates a pulsating fluid jet as a precise, minimally invasive and cold technique for bone cement removal. We utilize the pulsating fluid jet device to remove bone cement from samples designed to mimic clinical conditions. The effectiveness of a novel in-house designed long nozzle was tested to enable minimally invasive procedures. Audio signal monitoring, complemented by our introduced novel data correlation algorithm S4D-Bio, was employed to address challenges like visibility obstruction from splashing. The experiments aim to evaluate the effectiveness of our novel in-house designed long nozzle for minimally invasive removal of bone cement using a pulsating fluid jet as well as the prediction accuracy of the erosion rate. Within our experiments, we generate a comprehensive dataset of erosion profiles and their equivalent audio signals and make it available open-source. The use of SSMs yields experimentally demonstrated precise control over the predictive erosion process with a prediction accuracy of 98.93%. The study also demonstrates, that the pulsating fluid jet device, coupled with advanced audio monitoring techniques, is a highly effective cyber-physical system for estimating erosion depth under controlled conditions. On the other hand, this study presents the first application of SSMs in pulsating fluidjet surgery technology, marking a significant novelty. This research introduces the components of a future system for minimally invasive, cold and adaptive bone cement removal in orthopedic applications.
BibTeX
@inproceedings{schaller2025audio,
  author = {Melanie Schaller and Sergej Hloch and Akash Nag and Dagmar Klichová and Frank Pude and Michal Zeleňák and Nick Janssen and Bodo Rosenhahn},
  title = {{Audio monitoring of bone cement disintegration in pulsating fluid jet surgery under laboratory conditions}},
  booktitle = {Results},
  year = {2025},
  doi = {10.1016/j.rineng.2025.108762},
}