Assessment of Bias for MRI Diffusion Tensor Imaging Using SIMEX

Carolyn B. Lauzon, Andrew J. Asman, Ciprian M. Crainiceanu, Brian C. Caffo, Bennett A. LandmanMICCAI 2011 · OpenReview

Abstract

Diffusion Tensor Imaging (DTI) is a Magnetic Resonance Imaging method for measuring water diffusion in vivo. One powerful DTI contrast is fractional anisotropy (FA). FA reflects the strength of water’s diffusion directional preference and is a primary metric for neuronal fiber tracking. As with other DTI contrasts, FA measurements are obscured by the well established presence of bias. DTI bias has been challenging to assess because it is a multivariable problem including SNR, six tensor parameters, and the DTI collection and processing method used. SIMEX is a modern statistical technique that estimates bias by tracking measurement error as a function of added noise. Here, we use SIMEX to assess bias in FA measurements and show the method provides; i) accurate FA bias estimates, ii) representation of FA bias that is data set specific and accessible to non-statisticians, and iii) a first time possibility for incorporation of bias into DTI data analysis.
BibTeX
@inproceedings{lauzon2011assessment,
  author = {Carolyn B. Lauzon and Andrew J. Asman and Ciprian M. Crainiceanu and Brian C. Caffo and Bennett A. Landman},
  title = {{Assessment of Bias for MRI Diffusion Tensor Imaging Using SIMEX}},
  booktitle = {International Conference on Medical Image Computing and Computer-Assisted Intervention},
  year = {2011},
  url = {https://openreview.net/forum?id=IKRLyy6UJW},
}

Loading related content...

Figure