Artificial Intelligence for Hemodynamic Monitoring with a Wearable Electrocardiogram Monitor

Daphne E. Schlesinger, Ridwan Alam, Roey Ringel, Eugene Pomerantsev, Srikanth Devireddy, Pinak Shah, Joseph Garasic, Collin M. StultzmedRxiv 2024

Abstract

Background: The ability to non-invasively measure left atrial pressure would facilitate the identification of patients at risk of pulmonary congestion and guide proactive heart failure care. Wearable cardiac monitors, which record single-lead electrocardiogram data, provide information that can be leveraged to infer left atrial pressures.

Methods: We developed a deep neural network using single-lead electrocardiogram data to determine when the left atrial pressure is elevated. The model was developed and internally evaluated using a cohort of 6,739 samples from the Massachusetts General Hospital (MGH) and externally validated on a cohort of 4,620 samples from a second institution. We then evaluated the model on patch-monitor electrocardiographic data on a small prospective cohort.

Results: The model achieves an area under the receiver operating characteristic curve (AUROC) of 0.80 for detecting elevated left atrial pressures on an internal holdout dataset from MGH and 0.76 on an external validation set from a second institution. Evaluation of the model on a prospective dataset of patch-monitor signals yielded an AUROC of 0.875 for identifying elevated pressures when acquired close to the time of catheterization.

Conclusions: These results demonstrate the utility and the potential of ambulatory cardiac hemodynamic monitoring with electrocardiogram patch-monitors.

Introduction

Heart failure (HF) management is a high-stakes balancing act. A critical metric for these patients is Left Atrial Pressure (LAP). When LAP rises, it signals "pulmonary congestion"—fluid backing up into the lungs—which often leads to emergency room visits and hospital readmissions. Currently, the gold standard for measuring this pressure is mean pulmonary capillary wedge pressure (mPCWP), obtained via Right Heart Catheterization (RHC). While highly accurate, RHC is an invasive procedure that cannot be performed in a home setting. Non-invasive alternatives, such as Doppler ultrasound, require highly skilled technicians and are difficult to perform outside a clinical environment.

To bridge this gap, researchers developed the Cardiac Hemodynamic AI monitoring System (CHAIS), designed to infer internal cardiac pressures from the ubiquitous, non-invasive electrocardiogram (ECG) signal—specifically the single-lead data provided by modern wearables.

Methods

Model Architecture

CHAIS utilizes a single-lead adaptation of a residual neural network originally designed for 12-lead ECGs. To adapt the architecture for single-lead data (Lead I), the researchers employed a "12x Replication" trick, where the single-lead signal is replicated 12 times to form the input tensor. This approach was found to perform better than using a single-channel input.

Training and Reliability

The model underwent transfer learning, pre-trained on a dataset of over 242,000 patients to predict standard ECG intervals (PR, QRS, QT) and heart rate. This allowed the model to learn fundamental features of cardiac electrical signals before specializing in hemodynamic pressure detection. Additionally, the researchers implemented a "trustworthiness score" based on Shannon Entropy. By measuring model uncertainty, the system can flag and filter out predictions that are likely to be unreliable.

Study Design

The study utilized three distinct datasets:

  1. Internal Development (MGH): 6,739 samples from Massachusetts General Hospital, matching ECGs to same-day invasive RHC procedures.
  2. External Validation (BWH): 4,620 samples from Brigham and Women’s Hospital to test institutional generalizability.
  3. Prospective Wearable Cohort: 83 patients at MGH who wore a QOCA Portable ECG 101 patch-monitor for approximately 24 hours prior to a scheduled catheterization. The model was tested on this data in a "zero-shot" manner, without additional fine-tuning.

Results

Diagnostic Accuracy

The model demonstrated robust performance across all cohorts for detecting mPCWP > 18 mmHg:

  • Internal Validation: AUROC of 0.80.
  • External Validation: AUROC of 0.76.
  • Prospective Patch Performance: When analyzing patch data acquired within 2 hours of the catheterization, the model reached an AUROC of 0.875.

Key Insights

  • The "Time-Delta" Effect: Model performance improved significantly as the ECG recording time approached the time of the invasive measurement. This suggests that the ECG captures dynamic, real-time hemodynamic shifts rather than just static structural changes.
  • Parity with Ultrasound: The model’s AUROC is comparable to or exceeds the E/e’ ratio, a standard but complex echocardiographic metric used to estimate wedge pressure.
  • Rule-Out Capability: In low-risk populations (10% prevalence), the model showed a Negative Predictive Value (NPV) of over 95%, making it an excellent tool for ruling out dangerous pressure elevations.

Discussion

CHAIS represents a significant step toward "invisible" hemodynamic monitoring. By turning a simple wearable patch into a window into the heart's internal pressures, this technology could allow patients to monitor their heart health from home, potentially preventing hospitalizations through early intervention.

However, several limitations remain. The prospective patch study was limited to 83 patients, and larger ambulatory trials are necessary to confirm these findings. Furthermore, while patch monitors provide clean signals, the "noisier" data from wrist-worn smartwatches during daily activity presents a different challenge. Future work will also aim to move beyond binary thresholds toward continuous pressure estimation to better track patient trajectories over time.

Summary updated

Abstract

Abstract Background The ability to non-invasively measure left atrial pressure would facilitate the identification of patients at risk of pulmonary congestion and guide proactive heart failure care. Wearable cardiac monitors, which record single-lead electrocardiogram data, provide information that can be leveraged to infer left atrial pressures. Methods We developed a deep neural network using single-lead electrocardiogram data to determine when the left atrial pressure is elevated. The model was developed and internally evaluated using a cohort of 6739 samples from the Massachusetts General Hospital (MGH) and externally validated on a cohort of 4620 samples from a second institution. We then evaluated model on patch-monitor electrocardiographic data on a small prospective cohort. Results The model achieves an area under the receiver operating characteristic curve of 0.80 for detecting elevated left atrial pressures on an internal holdout dataset from MGH and 0.76 on an external validation set from a second institution. A further prospective dataset was obtained using single-lead electrocardiogram data with a patch-monitor from patients who underwent right heart catheterization at MGH. Evaluation of the model on this dataset yielded an area under the receiver operating characteristic curve of 0.875 for identifying elevated left atrial pressures for electrocardiogram signals acquired close to the time of the right heart catheterization procedure. Conclusions These results demonstrate the utility and the potential of ambulatory cardiac hemodynamic monitoring with electrocardiogram patch-monitors. Plain Language Summary Heart failure is a prevalent disorder that is challenging to manage. Part of what makes heart failure management challenging is that the onset of symptoms can be insidious as there are few robust tools that a clinical can leverage to estimate when a patient is likely to experience an episode of acute heart failure. While elevated intracardiac pressures are a reliable harbinger of worsening heart failure, these pressures are typically measured using an invasive, gold-standard approach, which can only be performed in an inpatient setting. A non-invasive method for detecting higher pressures inside of heart would be especially helpful for identifying worsening heart failure in an expeditious manner in the home environment. For this reason, we developed an artificial intelligence model to detect elevated pressures inside the heart using a non-invasive signal, the electrocardiogram (ECG, or EKG), which can be acquired from a wearable patch monitor device. Our results demonstrate that the model provides a reliable platform for the non-invasive assessment of cardiac pressures using data that can be obtained in the outpatient setting.
BibTeX
@misc{schlesinger2024artificial,
  author = {Daphne E. Schlesinger and Ridwan Alam and Roey Ringel and Eugene Pomerantsev and Srikanth Devireddy and Pinak Shah and Joseph Garasic and Collin M. Stultz},
  title = {{Artificial Intelligence for Hemodynamic Monitoring with a Wearable Electrocardiogram Monitor}},
  year = {2024},
  doi = {10.1101/2024.04.01.24304487},
  howpublished = {medRxiv},
}
Figure