Application of slice-accelerated one-minute TGSE-BLADE DWI in clinical practices
Abstract
TGSE-BLADE DWI has been reported to reduce geometric distortion and susceptibility artifacts, however the long acquisition time prevent its clinical application. We reduced acquisition time to 1 minute for TGSE-BLADE DWI using slice acceleration (1-min TGSE-BLADE DWI). We compared distortion and artifacts between SS-EPI DWI and 1-min TGSE-BLADE DWI, and evaluated diagnostic performance for acute infarctions of 1-min TGSE-BLADE DWI. The result shows that the 1-min TGSE-BLADE-DWI has better quality image in terms of distortion and artifacts, and higher diagnostic performance for acute infarctions. TGSE-BLADE DWI with slices acceleration is a promising method for evaluating lesions in acute stroke patients.
BibTeX
@inproceedings{okuchi2024application,
author = {Sachi Okuchi and Yasutaka Fushimi and Satoshi Nakajima and Akihiko Sakata and Takuya Hinoda and Sayo Otani and Azusa Sakurama and Krishna Pandu Wicaksono and Hiroshi Tagawa and Yan Wang and Satoshi Ikeda and Shuichi Ito and Miyuki Takiya and Kun Zhou and Yuji Nakamoto},
title = {{Application of slice-accelerated one-minute TGSE-BLADE DWI in clinical practices}},
booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
year = {2024},
doi = {10.58530/2023/2083},