Animal experiment on osseointegration of porous titanium root analogue implants with composite CSn-TAK242 coating

Hui Li, Dan Luo, Yudong Gao, Dashan Wang, Jianjun Yang, Zexian XuFrontiers in Bioengineering and Biotechnology 2025 · DOI

Abstract

Porous titanium root analogue implants consistent with the target root morphology were successfully fabricated using digital medical technology and 3D printing. The composite CSn-TAK242 coating further enhanced the osseointegration effects of these implants.

Introduction

Standard dental implants are typically cylindrical or conical, which often fails to match the complex anatomy of a natural tooth socket, especially in multi-rooted molar regions. This anatomical mismatch can lead to poor initial stability and "stress shielding," a phenomenon where the stiff metal implant carries all the mechanical load, causing the surrounding bone to weaken and resorb over time.

Furthermore, the act of implantation triggers an inflammatory response. While some inflammation is necessary for healing, excessive activation of the Toll-like receptor 4 (TLR4) pathway can lead to bone loss and implant failure. This research explores a dual-action solution: using 3D printing to ensure a precise mechanical fit and a bioactive coating to manage the biological environment.

Materials and Methods

The researchers combined digital dentistry with advanced material science to create Root Analogue Implants (RAIs):

  1. Digital Modeling & 3D Printing: Using CBCT scans of beagle teeth, the team created 3D models of the roots. They then used Selective Laser Melting (SLM) to print these shapes using medical-grade titanium powder ($Ti_6Al_4V$), incorporating a porous, bone-like trabecular structure with 60%–75% porosity.
  2. Bioactive Coating: A composite coating was developed using chitosan nanoparticles (CSn) loaded with TAK242, a small molecule that specifically inhibits TLR4 signaling.
  3. Experimental Model: 48 implants were placed into the extraction sockets of six beagles, divided into three groups: a bare titanium control, a CSn-only group, and the CSn-TAK242 experimental group.
Digital modeling and fabrication process of 3D-printed porous titanium root analogue implants.

Key Results

Superior Bone Integration

The Bone-to-Implant Contact (BIC)—a gold standard for implant success—was significantly higher in the CSn-TAK242 group (78.6%) compared to the control group (68.11%). Micro-CT imaging further showed that the CSn-TAK242 group had significantly higher bone volume/total volume (BV/TV) and thicker trabeculae (Tb.Th) than the other groups.

Reduced Inflammation

RT-qPCR confirmed that the expression of MyD88 and TRIF—key downstream signals of the inflammatory TLR4 pathway—was significantly lower in the group receiving the drug-loaded coating. Interestingly, while the drug inhibited the action of TLR4, the body responded by increasing the mRNA expression of TLR4, likely a compensatory feedback mechanism that did not negate the drug's effectiveness.

Mechanical Stability

All groups showed good initial stability, with Implant Stability Quotient (ISQ) values ranging from 62 to 64. This proves that the 3D-printed shape alone provides a solid mechanical foundation without needing bone grafts or extensive socket preparation.

Discussion and Conclusion

The study confirms that morphology matters; 3D-printed RAIs achieve high initial stability because they fit the extraction socket perfectly. The porous design not only lowers the stiffness of the titanium to match bone but also provides a scaffold for "micromechanical interlocking," where bone grows directly into the implant.

By inhibiting the TLR4 pathway, the CSn-TAK242 coating shifts the biological environment toward bone formation (osteogenesis) rather than bone destruction (resorption). While the study noted a failure rate of approximately 14.5% due to surgical complexity and postoperative care challenges in a canine model, the results provide a strong foundation for future human clinical trials. Future work will focus on optimizing surgical placement and conducting longer-term studies to evaluate mechanical loading over time.

Summary updated

Abstract

Porous titanium root analogue implants consistent with the target root morphology were successfully fabricated using digital medical technology and 3D printing. The composite CSn-TAK242 coating further enhanced the osseointegration effects of these implants.
BibTeX
@inproceedings{li2025animal,
  author = {Hui Li and Dan Luo and Yudong Gao and Dashan Wang and Jianjun Yang and Zexian Xu},
  title = {{Animal experiment on osseointegration of porous titanium root analogue implants with composite CSn-TAK242 coating}},
  booktitle = {Frontiers in Bioengineering and Biotechnology},
  year = {2025},
  doi = {10.3389/fbioe.2025.1673758},
}
Figure