The Jiangmen Underground Neutrino Observatory (JUNO) is a next-generation neutrino experiment designed to determine the neutrino mass hierarchy. Central to its success is a 20-kiloton liquid scintillator detector using Linear Alkylbenzene (LAB) as the solvent. To achieve an unprecedented energy resolution of $\sigma_E/E \leqslant 3\%$, the LAB must exhibit exceptional transparency in the 350–450 nm wavelength range. As industrial production of LAB has improved, its transparency has approached the measurement limits of existing apparatuses, necessitating significant upgrades to hardware and analytical models. This paper details the development of a high-precision measurement system and a comprehensive error analysis framework.
Experimental Challenges and Apparatus Upgrades
Measuring light attenuation lengths exceeding 25 meters is technically demanding. In such ultra-transparent media, the difference in light intensity over short distances is minimal, making measurements highly sensitive to stray light, beam divergence, and sensor noise.
To overcome these hurdles, the research team implemented two primary hardware improvements:
- Collimated Fiber-Coupled Source: The original LED source had a wide $6^\circ$ divergence angle, leading to significant aberrations and stray light. The upgraded system utilizes a multi-mode optical fiber coupled with a precision collimator. This configuration produces a much tighter, parallel beam, significantly reducing the "halo" effect and ensuring light interacts more cleanly with the sample.
- High-Performance PMT: The detection system was upgraded from a Hamamatsu CR135 to an R7724 photomultiplier tube (PMT). The new PMT offers superior time response and lower noise characteristics, which are critical for distinguishing subtle variations in light intensity.
Advanced Modeling and Error Analysis
Beyond hardware, the team refined the mathematical treatment of the data. While the Beer-Lambert law is the standard for absorption, it fails to account for Rayleigh scattering, which becomes dominant in high-purity LAB.
- Rayleigh Scattering Correction: A modified model was introduced to account for scattered light that remains within the PMT's field of view. By incorporating a geometry-dependent term ($\mu_{mod}$), the researchers achieved a significantly better fit to the experimental data, reducing systematic biases.
- Monte Carlo Simulations: To rigorously quantify uncertainties, the team employed Monte Carlo methods. By simulating 100,000 measurement cycles, they could accurately estimate the impact of fluctuations in light intensity and liquid level measurements on the final attenuation length.
Key Results and Impact
The study evaluated several batches of improved LAB (NJ64–NJ68). The most significant result was obtained for the NJ66 sample, which reached a record light attenuation length of $29.90 \pm 0.95$ m. This value represents a new benchmark for LAB transparency and provides strong evidence that JUNO can meet its stringent energy resolution requirements.
The upgrades pushed the theoretical measurement limit ($L_{max}$) of the apparatus from approximately 29 meters to over 80 meters. Validation through independent measurements confirmed the stability and reproducibility of the new system.
Conclusion and Future Outlook
The successful measurement of record-breaking transparency in LAB validates the ongoing collaboration between JUNO researchers and industrial partners. While the current upgrades have significantly reduced errors, future work will focus on further refining the scattering models to account for diffuse reflections and the finite spectral width of the LED source. These refinements will ensure that JUNO remains at the forefront of neutrino physics.
Summary updated