An optimized sliding rail-assisted micrometer system for sensing volume measurement of open-ended coaxial probes in breast cancer dielectric property analysis

Shuai Chen, Zhongliang Lu, Qiang Huang, Gaowei Zhao, Sun Zheng, Zhou Jing, Yupeng LiaoFrontiers in Bioengineering and Biotechnology 2025 · DOI

Abstract

The results indicate that both the sensing depth and radius of the probe increase significantly with larger coaxial probe aperture sizes. Furthermore, a smaller aperture reduces the influence of tissue heterogeneity on measurements, while the effective sensing volume remains consistent across frequencies.

Introduction: The Challenge of Tissue Heterogeneity

In breast cancer diagnostics, measuring the dielectric properties of tissue—how it interacts with electromagnetic fields—is a promising technique. Cancerous tissues typically exhibit higher water content and altered membrane permeability, creating distinct dielectric signatures compared to healthy fat or glandular tissue. The Open-Ended Coaxial Probe (OECP) is the standard tool for these measurements due to its non-destructive nature. However, its accuracy is often compromised by tissue heterogeneity. Because biological tissue is a complex mixture, the "sensing volume" of a probe may overlap multiple tissue types, leading to averaged data and potential misdiagnosis. Understanding the exact spatial boundaries of what a probe "sees" is critical for clinical precision.

Methods: Precision Mapping with Micrometer Accuracy

To address the limitations of existing measurement systems, researchers developed an optimized sliding rail-assisted micrometer system. This setup achieves a resolution of 0.01 mm, a tenfold improvement over standard methods, by integrating high-precision ball linear slides with digital micrometer heads.

The experimental setup utilizes a Vector Network Analyzer (VNA) to measure the reflection coefficient at the probe tip. To simulate the transition between different tissue types, a "bilayer model" was employed. A probe was placed against a Teflon block (simulating low-permittivity fat) and moved incrementally through various Liquids Under Test (LUT), such as ethanol and deionized water (simulating high-permittivity tumor tissue).

Figure 1: VNA measurement setup and calibration procedure, showing the integration of the probe, analyzer, and control console.

Validation and Key Results

The system's accuracy was validated using reference materials and biological samples, including porcine tissue and human breast cancer samples. A key finding was the stability of the sensing volume across the 300 MHz to 1.5 GHz frequency range. This frequency independence simplifies clinical application, as spatial expectations remain constant regardless of the operating frequency.

Figure 6: Comparison of measured and standardized conductivity for ethanol, methanol, and normal saline, demonstrating the system's high measurement accuracy.

The study quantified the spatial boundaries for two common probe diameters:

  • 2.20 mm Probe: Sensing depth of 0.44–0.62 mm and radius of 0.36–0.63 mm.
  • 3.58 mm Probe: Sensing depth of 0.75–0.98 mm and radius of 0.71–0.99 mm.

Larger apertures provide a deeper view but are more susceptible to errors from heterogeneous layers. For early-stage tumor detection, the 2.20 mm probe is preferred due to its smaller footprint, which minimizes interference from surrounding healthy tissue.

Conclusion and Future Directions

While the sliding rail system offers superior mechanical stability and eliminates electromagnetic interference compared to motorized stages, challenges remain. The "air gap" problem—where imperfect contact between the probe and tissue distorts data—requires further mechanical refinement. Future research will also explore more complex 3D tissue phantoms incorporating particles to simulate microscopic heterogeneities like microcalcifications.

Summary updated

Abstract

The results indicate that both the sensing depth and radius of the probe increase significantly with larger coaxial probe aperture sizes. Furthermore, a smaller aperture reduces the influence of tissue heterogeneity on measurements, while the effective sensing volume remains consistent across frequencies.
BibTeX
@inproceedings{chen2025optimized,
  author = {Shuai Chen and Zhongliang Lu and Qiang Huang and Gaowei Zhao and Sun Zheng and Zhou Jing and Yupeng Liao},
  title = {{An optimized sliding rail-assisted micrometer system for sensing volume measurement of open-ended coaxial probes in breast cancer dielectric property analysis}},
  booktitle = {Frontiers in Bioengineering and Biotechnology},
  year = {2025},
  doi = {10.3389/fbioe.2025.1575142},
}
Figure