An MR technique for simultaneous quantitative imaging of water content, conductivity and susceptibility, with application to brain tumours using a 3T hybrid MR-PET scanner

Yupeng Liao, Ana‐Maria Oros‐Peusquens, Johannes Lindemeyer, Nazim Lechea, Carolin Weiß -Lucas, Karl‐Josef Langen, N. Jon ShahScientific 2019 · DOI

TL;DR

This paper introduces a novel 2D multi-echo gradient echo MRI technique that simultaneously quantifies water content, electrical conductivity, and magnetic susceptibility in the brain. This method significantly reduces acquisition time compared to independent measurements, offering a more comprehensive and efficient approach for characterizing brain tissue, particularly in the context of brain tumors.

What Problem Does This Paper Address?

Quantitative Magnetic Resonance Imaging (qMRI) offers valuable insights into tissue physiology by measuring parameters like water content, electrical conductivity, and magnetic susceptibility. However, current methods for mapping these properties are typically performed independently, leading to long acquisition times and potential co-registration issues. This fragmentation hinders the comprehensive understanding of tissue states, especially in complex pathologies like brain tumors where alterations in these properties can indicate disease progression and treatment response. The paper aims to bridge this gap by developing a single, efficient MRI sequence that can simultaneously acquire all three quantitative maps.

Methods

The core of this work is a novel post-processing technique applied to data from a 2D multi-echo gradient echo (GRE) sequence. This technique leverages both the magnitude and phase information across multiple echo times to derive the three quantitative parameters:

  • Water Content: Derived from the magnitude signal decay, after applying T2* and transmit-receive (B1) field bias corrections. The signal intensity at zero echo time (S0) is proportional to the proton density, which is then normalized to cerebrospinal fluid (CSF) as a reference for 100% water content.
  • Magnetic Susceptibility: Calculated from the phase evolution during the multi-echo train. This involves unwrapping the phase, removing background fields (using a method called MUBAFIRE), and then reconstructing the susceptibility map using a dipole inversion algorithm.
  • Electrical Conductivity: Retrieved from the phase value extrapolated to zero echo time (TE=0). This is achieved by fitting the phase evolution and then applying a phase-based magnetic resonance electrical properties tomography (MREPT) method, specifically using the Laplacian of the transmit field phase.

The method was validated using phantoms and then applied to healthy volunteers and brain tumor patients on a 3T hybrid MR-PET scanner.

Datasets

The study utilized data from:

  • Phantoms: A multi-compartment phantom with varying concentrations of D2O (for water content), NaCl (for conductivity), and Gadolinium (for susceptibility) was used for initial validation.
  • Healthy Volunteers: Two healthy volunteers were scanned.
  • Brain Tumor Patients: Three patients with different types of brain tumors (meningioma, glioblastoma, and metastasis) were scanned.

The paper also references and compares results with conventional 3D GRE-based protocols and clinical MR sequences (MP-RAGE, FLAIR) and O-(2-[18F]fluoroethyl)-L-tyrosine (FET)-PET for tumor delineation.

Key Insights and Ideas

  • Simultaneous Quantification is Key: The ability to acquire water content, conductivity, and susceptibility maps in a single scan is a significant advancement, reducing scan time and eliminating co-registration challenges.
  • Leveraging Magnitude and Phase: The method effectively utilizes both magnitude decay (for water content) and phase evolution (for susceptibility and conductivity) from the GRE signal, extracting more information than previous approaches.
  • Tissue Viability Insights: The combined information from these three parameters offers a more nuanced understanding of tissue health and pathology than any single parameter alone, potentially aiding in tumor characterization and treatment monitoring.
  • Tumor Heterogeneity: The study highlights that different tumor types and even different regions within a tumor can exhibit distinct profiles across these quantitative parameters, suggesting potential for more precise diagnosis and prognosis.

Key Results

  • The proposed 2D multi-echo GRE method demonstrated comparable results to reference 3D protocols for phantoms and healthy volunteers, with good agreement in water content, conductivity, and susceptibility maps.
  • In brain tumor patients, the method successfully visualized abnormal regions, showing increased water content in tumors and surrounding edema, consistent with clinical observations.
  • Electrical conductivity maps revealed increased values in abnormal tissues, potentially linked to ion concentration changes.
  • Magnetic susceptibility showed more variable contrasts, with some tumors exhibiting paramagnetic and others diamagnetic characteristics, suggesting complex underlying tissue changes.
  • The acquisition time for the quantitative mapping sequence was significantly reduced (e.g., 7 min 21 s for patients) compared to traditional methods.

Limitations and Future Work

  • Slice-Staggering Effect: The 2D nature of the acquisition can lead to slice-staggering artifacts, particularly affecting susceptibility maps, which could be mitigated by advanced 2D phase processing techniques.
  • Accuracy of TE=0 Extrapolation: The accuracy of conductivity mapping relies on the precise extrapolation of phase to TE=0. In regions with very short T2* values, later echoes may fall below the noise floor, impacting this extrapolation.
  • Potential for Faster Acquisition: The authors suggest that further reductions in TR (e.g., to 5s) and flip angle, combined with noise reduction techniques like Principal Component Analysis (PCA), could enable whole-brain multi-parametric mapping in under 5 minutes.
  • Further Validation: While promising, the study acknowledges the need for further studies with larger patient cohorts to confirm findings regarding the differences between malignant and benign tumors.
  • Investigating Other Pathologies: The authors suggest that future work could explore the application of this method to other neurological conditions beyond brain tumors.

Summary updated

Abstract

Approaches for the quantitative mapping of water content, electrical conductivity and susceptibility have been developed independently. The purpose of this study is to develop a method for simultaneously acquiring quantitative water content, electrical conductivity and susceptibility maps based on a 2D multi-echo gradient echo sequence. Another purpose is to investigate the changes in these properties caused by brain tumours. This was done using a 3T hybrid magnetic resonance imaging and positron emission tomography (MR-PET) scanner. Water content maps were derived after performing T<sub>2</sub>* and transmit-receive field bias corrections to magnitude images essentially reflecting only the H<sub>2</sub>O content contrast. Phase evolution during the multi-echo train was used to generate field maps and derive quantitative susceptibility, while the conductivity maps were retrieved from the phase value at zero echo time. Performance of the method is demonstrated on phantoms and two healthy volunteers. In addition, the method was applied to three patients with brain tumours and a comparison to maps obtained from PET using O-(2-[18 F]fluoroethyl)-L-tyrosine and clinical MR images is presented. The combined information of the water content, conductivity and susceptibility may provide additional information about the tissue viability. Future studies can benefit from the evaluation of these contrasts with shortened acquisition times.