An IMU-Based Real-Time Measuring System for Acetabular Prosthesis Implant Angles in THR Surgeries

Hong Chen, Zhecheng Yang, Jilin Zhang, Jie Liu, Hao Tang, Yixin Zhou, Binjie Zhu, Zhihua WangIEEE Sensors Journal 2021 · DOI

Abstract

With the aging of the population, the number of total hip replacement (THR) surgeries increases year by year. To improve the success rate of THR operations, the measurement of acetabular prosthesis angles should be precise during surgery. We design an inertial measurement unit (IMU) based real-time measurement system, which measures the initial attitude of the pelvis coordinate and traces the real-time attitude of the pelvis coordinate and prosthesis axis. By collecting real-time data of different pelvic and prosthesis attitudes in quaternion representation through accelerometers and gyroscopes, the relative position between the pelvis and prosthesis is calculated in real time. An optimized algorithm is proposed to obtain prosthesis implant angles accurately. Experimental results show that the measured implant angles agree with those calculated by radiographic values, with root mean square errors (RMSE) less than 3.52 degrees, within the safe range of THR requirements. The system can trace real-time dislocation of the prosthesis during THR surgeries.

Introduction

In Total Hip Replacement (THR) surgery, the orientation of the acetabular cup is paramount. Misalignment can lead to impingement, accelerated wear, and most critically, post-operative dislocation—the leading cause of THR failure. Surgeons traditionally rely on mechanical guides or visual estimation, which are prone to significant error. While computer-assisted navigation systems (CANS) using CT scans or optical tracking offer high precision, they are often hindered by high costs, radiation exposure, and complex line-of-sight requirements in a crowded operating room.

Inertial Measurement Units (IMUs) present a promising alternative, offering a low-cost, radiation-free, and compact solution. However, standard IMU systems face two major hurdles: magnetic interference from surgical equipment and "gimbal lock" singularities when using Euler angles. This research introduces a robust, 6-axis IMU system designed specifically for the rigors of the surgical environment.

System Architecture and Methodology

The developed system comprises three specialized instruments, each housing a 6-axis IMU consisting of a 3-axis accelerometer and a 3-axis gyroscope. By intentionally omitting magnetometers, the system remains immune to the magnetic distortions caused by metal surgical tables and power tools.

The three specialized measurement instruments: (a) IAMI for initial pelvic attitude, (b) RAMI for real-time pelvic tracking, and (c) PAMI for prosthesis orientation.

The workflow involves three key components:

  1. IAMI (Initial Attitude Measurement Instrument): This device is used at the beginning of the procedure to define the patient's Anterior Pelvic Plane (APP).
  2. RAMI (Real-time Attitude Measurement Instrument): This module is firmly fixed to the pelvis, allowing the system to track pelvic movement throughout the surgery.
  3. PAMI (Prosthesis Attitude Measurement Instrument): Attached to the prosthesis impactor handle, it tracks the real-time orientation of the implant.

A critical innovation is the use of quaternion-based data fusion combined with an Extended Kalman Filter (EKF). Unlike traditional Euler angles (roll, pitch, yaw), quaternions do not suffer from gimbal lock, a mathematical state where a degree of freedom is lost at certain orientations (e.g., ±90° pitch). This ensures the system remains reliable regardless of the patient's posture or the angle of approach.

Clinical Context and Angle Definition

The system's primary output is the measurement of the Radiographic Inclination (RI) and Radiographic Anteversion (RA) angles. These are defined relative to the pelvic coordinate system, which is established using the APP—a plane defined by the two anterior superior iliac spines and the pubic symphysis.

Geometric definition of the pelvic coordinate system and the target implant angles (RI and RA) relative to the Anterior Pelvic Plane (APP).

By tracking the RAMI and PAMI simultaneously, the system calculates the relative orientation of the prosthesis to the pelvis. This "dynamic referencing" means that if the patient shifts on the operating table after the initial calibration, the measurements remain valid, solving a common practical headache for surgeons.

Experimental Evaluation

The system was rigorously tested using a physical pelvis model and a custom-built protractor device to provide ground truth.

  • Fixed Angle Trials: 34 experiments were conducted where the prosthesis was held at constant angles. The results were compared against radiographic values obtained via CT scans, the current "gold standard." The system achieved an RMSE of 3.18° for RA and 3.51° for RI.
  • Variable Angle Trials: To test dynamic tracking, 41 experiments were performed where implant angles were varied across a wide range (-40° to 40°). The system maintained an RMSE of less than 3.27°.
  • Operational Performance: The IMU sensors showed a maximal drift of only 1.5° over 5 minutes, well within the time required for cup placement. Furthermore, the system operates with a latency of just 2.5ms, providing instantaneous feedback.

Conclusion

The developed IMU-based system offers a low-cost, accurate, and radiation-free alternative for THR navigation. By utilizing a 6-axis, magnetometer-free approach and quaternion mathematics, it successfully addresses the limitations of current surgical tools. Future work will focus on clinical trials to validate the system in human subjects and account for soft tissue dynamics.

Summary updated

Abstract

With the aging of population, the number of total hip replacement (THR) surgeries increases year by year. To improve the success rate of THR operation, the measurement of acetabular prosthesis angles should be precise during the surgery. We design an inertial measurement unit based real-time measurement system, which measures the initial attitude of the pelvis coordinate, and traces the real-time attitude of the pelvis coordinate and prosthesis axis. By collecting real-time data of diff- erent pelvic and prosthesis attitudes in quaternion representation through accelerometers and gyroscopes, the relative position between pelvis and prosthesis is calculated in real time. An optimized algorithm is proposed to obtain prosthesis implant angles accurately from the data. The experimental results show that the measured implant angles agree with that calculated by radiographic value, the root mean square errors (RMSE) are less than 3.52 degrees, which is within the safe range of THR requirement. Furthermore, no matter how the implant angles change, the measured values agree with reference values, the RMSE is less than 3.27 degrees. The experimental results show that our system can trace the real-time dislocation of the prosthesis during the THR surgeries.
BibTeX
@article{chen2021imu,
  author = {Hong Chen and Zhecheng Yang and Jilin Zhang and Jie Liu and Hao Tang and Yixin Zhou and Binjie Zhu and Zhihua Wang},
  title = {{An IMU-Based Real-Time Measuring System for Acetabular Prosthesis Implant Angles in THR Surgeries}},
  journal = {IEEE Sensors Journal},
  year = {2021},
  doi = {10.1109/jsen.2021.3091583},
}
Figure