An Automatic Model-based System for Joint Space Measurements on Hand Radiographs: Initial Experience

Philipp Peloschek, Georg Langs, Michael Weber, Johannes Sailer, Michael Reisegger, H. Imhof, Horst Bischof, F. KainbergerRadiology 2007 · DOI

Abstract

This ethics committee-approved pilot study was performed with informed consent. A Web-based service that was developed for automated measurement of joint space and automatic analysis of radiographs was tested prospectively. A total of 160 metacarpophalangeal joint spaces were measured in 20 patients (average age, 48 years; age range, 18-89 years; 16 women) suspected of having rheumatoid arthritis or osteoarthritis. The technical success rate was 93%. The smallest detectable difference in repeated automatic joint space width measurements varied from 0.08 to 0.31 mm, and the coefficient of variation was 2%-7%. Compared with the reference standard (interactive segmentation of the joint space widths) measurements, results were within a mean error of 0.19-0.40 mm. The proposed Web-based service enables reproducible joint space measurements to be obtained in metacarpophalangeal joints with moderate erosive and osteophytic disease.
BibTeX
@inproceedings{peloschek2007automatic,
  author = {Philipp Peloschek and Georg Langs and Michael Weber and Johannes Sailer and Michael Reisegger and H. Imhof and Horst Bischof and F. Kainberger},
  title = {{An Automatic Model-based System for Joint Space Measurements on Hand Radiographs: Initial Experience}},
  booktitle = {Radiology},
  year = {2007},
  doi = {10.1148/radiol.2452061281},
}