'Uniclear' water-based brain clearing for light sheet imaging of electrode tracks v1

Susu Chen, Karel Svoboda2019 · DOI

Abstract

The 'Uniclear' protocol provides a streamlined, water-based method for clearing whole mouse brains, rendering them transparent for light-sheet fluorescence microscopy (LSFM). By focusing on mechanical robustness and precise refractive index matching, this technique enables the 3D mapping of electrode tracks and anatomical features without the use of toxic organic solvents. The resulting specimens are structurally stiff enough for vertical mounting and high-resolution imaging.

Introduction

In neuroscience, understanding the exact placement of electrodes or probes post-experiment is crucial for interpreting physiological data. However, the mammalian brain is naturally opaque due to high lipid content in cell membranes and myelin, which scatters light. Traditional clearing methods often rely on aggressive organic solvents (such as BABB or iDISCO) which, while effective, can be toxic, cause significant tissue shrinkage, or quench fluorescent signals. Furthermore, many methods leave tissue fragile and difficult to handle. 'Uniclear' was developed to provide a safer, water-based alternative that preserves structural integrity while achieving the high transparency needed for light-sheet imaging.

Methodology

The Uniclear protocol is a multi-step chemical process designed to remove light-scattering lipids and match the refractive index (RI) of the tissue to its surrounding medium.

  1. Fixation: The brain is fixed using 4% Paraformaldehyde (PFA).
  2. Delipidation (SBiP Buffer): The brain is immersed in a specialized "SBiP" buffer—a mixture of Sodium Dodecyl Sulfate (SDS), 2-methyl-2-butanol, and 2-propanol. The sample is incubated at 37°C with daily buffer changes for 14 days to wash away light-scattering lipids.
  3. Refractive Index (RI) Matching: Following delipidation, the brain is transitioned through increasing concentrations of an OptiPrep (iodixanol) and Nycodenz solution. By matching the RI of the liquid to the protein-rich tissue scaffold, the brain becomes effectively transparent.
  4. Imaging: Cleared specimens are imaged using a light-sheet microscope (specifically the Zeiss Z1).

Key Results and Insights

The Uniclear protocol successfully renders adult mouse brains transparent enough for full-depth imaging. A standout feature of the method is the mechanical robustness of the resulting specimen; unlike the "jelly-like" brains produced by some protocols, Uniclear maintains enough structural stiffness to be mounted vertically for light-sheet imaging without deforming.

Figure 3: A cleared mouse brain imaged to show internal structural details and transparency achieved via the Uniclear protocol.

Key performance highlights include:

  • Imaging Resolution: The authors demonstrate imaging with a voxel size of 1.22 µm x 1.22 µm x 8 µm using a 5x objective.
  • Track Localization: The method is specifically optimized for visualizing electrode tracks. By capturing tissue autofluorescence at 488 nm and CM-DiI labeled tracks at 561 nm, researchers can precisely map recording sites in 3D space.
  • Safety and Compatibility: By avoiding harsh solvents like dibenzyl ether (DBE), the protocol is safer for researchers and potentially more compatible with a wider range of endogenous fluorescent labels.

Discussion and Limitations

While Uniclear provides high-quality clearing and structural stability, the 14-day delipidation period is relatively slow compared to some rapid clearing techniques, which may limit throughput for large-scale studies. Additionally, the protocol is specifically optimized for light-sheet microscopy; imaging on standard confocal systems would still be limited by objective working distances. Future iterations may explore ways to accelerate delipidation without sacrificing the mechanical robustness that makes this protocol unique.

Summary updated

Abstract

UniClear procedure for whole mouse brain clearing and refractive index matching. The advantages of this method are that the method is water-based and produces a mechanically robust specimen. Cleared brains can be imaged with lightsheet microscopes (e.g. Zeiss Z1 lightsheet).
BibTeX
@misc{chen2019amp,
  author = {Susu Chen and Karel Svoboda},
  title = {{\&\#39;Uniclear\&\#39; water-based brain clearing for light sheet imaging of electrode tracks v1}},
  year = {2019},
  doi = {10.17504/protocols.io.zndf5a6},
}
Figure