Advanced Photonic Sensors Based on Interband Cascade Lasers for Real-Time Mouse Breath Analysis

Erhan Tütüncü, Markus Nägele, S. Becker, M. Fischer, Johannes Koeth, Christian Wolf, Stefan Köstler, Volker Ribitsch, Andrea Teuber, Michael Gröger, Sandra Kress, Martin Wepler, Ulrich Wachter, J. Vogt, Peter Radermacher, Boris MizaikoffACS 2018 · DOI

Abstract

A multiparameter gas sensor based on distributed feedback interband cascade lasers emitting at 4.35 μm and ultrafast electro-spun luminescence oxygen sensors has been developed for the quantification and continuous monitoring of <sup>13</sup>CO<sub>2</sub>/<sup>12</sup>CO<sub>2</sub> isotopic ratio changes and oxygen in exhaled mouse breath samples. Mid-infrared absorption spectra for quantitatively monitoring the enrichment of <sup>13</sup>CO<sub>2</sub> levels were recorded in a miniaturized dual-channel substrate-integrated hollow waveguide using balanced ratiometric detection, whereas luminescence quenching was used for synchronously detecting exhaled oxygen levels. Allan variance analysis verified a CO<sub>2</sub> measurement precision of 1.6‰ during a 480 s integration time. Routine online monitoring of exhaled mouse breath was performed in 14 mechanically ventilated and instrumented mice and demonstrated the feasibility of online isotope-selective exhaled breath analysis within microliters of probed gas samples using the reported combined sensor platform.
BibTeX
@inproceedings{tutuncu2018advanced,
  author = {Erhan Tütüncü and Markus Nägele and S. Becker and M. Fischer and Johannes Koeth and Christian Wolf and Stefan Köstler and Volker Ribitsch and Andrea Teuber and Michael Gröger and Sandra Kress and Martin Wepler and Ulrich Wachter and J. Vogt and Peter Radermacher and Boris Mizaikoff},
  title = {{Advanced Photonic Sensors Based on Interband Cascade Lasers for Real-Time Mouse Breath Analysis}},
  booktitle = {ACS},
  year = {2018},
  doi = {10.1021/acssensors.8b00477},
}

Loading related content...

Figure