NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

Adjustable-Stiffness Hip Exoskeleton with Flexible Energy-Storage Module for 3D Gait Correction

Tianyu Xu, Zhenkun Sun, Sujiao Li, Hongyan Tang, Yanbin Zhang, Raymond Kai‐Yu Tong, Qiaoling Meng, Hongliu YuMachines 2025

TL;DR

This paper introduces a novel hip exoskeleton with an adjustable-stiffness energy-storage module designed for 3D gait correction. By dynamically adjusting stiffness based on gait phase and user parameters, the system significantly reduces joint variability and improves muscle activation, offering a promising solution for gait rehabilitation.

What Problem Does This Paper Address?

Current lower-limb exoskeletons often struggle to effectively address the diverse and complex gait abnormalities seen in individuals with neuromuscular or musculoskeletal disorders. Many existing devices are rigid, limiting natural joint movement and user comfort, and are primarily designed for sagittal plane motion, neglecting crucial multi-planar dynamics. Furthermore, many rely on passive support or simple phase-based control, lacking the ability to actively correct pathological gait patterns in real-time and adapt to individual user needs and varying walking speeds. This gap hinders the development of truly personalized and effective rehabilitation technologies.

Methods

The core innovation lies in the adjustable-stiffness flexible energy-storage module integrated into a hip exoskeleton. This module utilizes a flexible, cable-driven design with elastic elements that store and release energy to assist hip flexion, abduction, and adduction. The system's mechanical design is modular and flexible, avoiding rigid joint alignment to better accommodate natural human movement.

For control, the researchers developed a stiffness-gain scheduled PID control strategy. This strategy dynamically adjusts the system's stiffness in real-time based on gait phase and user-specific parameters, such as walking speed. The "stiffness gain" is a dynamic parameter that increases with walking speed to provide more support, and decreases at slower speeds for greater compliance. This allows for personalized assistance and adaptation to varying gait patterns.

A 3D gait correction model was constructed based on biomechanical analysis of the hip joint, targeting impairments in flexion, abduction, and adduction. This model informs the control strategy, enabling it to apply corrective forces throughout the gait cycle.

Datasets

The study utilized the GaitMotion multi-task gait dataset developed by the University of British Columbia as a reference for analyzing normal and hemiparetic circumduction gait.

Key Insights and Ideas

  • Flexible Design is Key: The move away from rigid structures towards a flexible, cable-driven design with elastic elements is crucial for accommodating natural joint movement and improving user comfort, a significant limitation in many existing exoskeletons.
  • Dynamic Stiffness for Personalized Assistance: The concept of a "stiffness-gain scheduled PID controller" that dynamically adjusts stiffness based on gait speed and phase is a powerful approach for providing truly personalized and adaptive gait assistance. This moves beyond static or pre-programmed assistance.
  • 3D Gait Correction is Essential: Addressing gait impairments in three dimensions (flexion, abduction, adduction) is vital for comprehensive rehabilitation, as many pathological gaits involve multi-planar deviations.
  • EMG as a Proxy for Muscle Load Redistribution: The observed changes in muscle activation (e.g., increased rectus femoris activity) provide strong evidence that the exoskeleton is not just assisting movement but actively redistributing muscle load, potentially reducing user effort and fatigue.

Key Results

  • Reduced Joint Variability: The exoskeleton demonstrated a significant reduction in joint trajectory variability, with a 22% decrease in the standard deviation of hip flexion angle. Knee and ankle joint deviations also decreased by 15% and 12%, respectively.
  • Improved Muscle Activation: EMG analysis showed a 21.4% increase in rectus femoris activity when wearing the exoskeleton, indicating enhanced muscle engagement and support. The gluteus medius showed a minimal increase (0.3%), while the long adductor muscle showed a 6.8% increase.
  • Enhanced Gait Symmetry and Stability: The results suggest improved mediolateral stability and gait balance, with more symmetrical hip abduction/adduction patterns and smoother flexion/extension transitions.

Limitations and Future Work

  • Small Sample Size: The study was conducted with a single subject, limiting the generalizability of the findings. A larger and more diverse participant group is needed to assess the system's adaptability across various conditions and gait abnormalities.
  • Portability and Comfort: The current reliance on external power sources and potential for discomfort during long-term use are noted limitations. Future work should focus on improving portability, battery life, and overall user comfort.
  • Real-World Performance: The system's performance in uncontrolled, real-world environments (e.g., varied terrain, dynamic movements) remains untested.
  • Long-Term Effects: Further investigation is required into the long-term effects of exoskeleton use on muscle adaptation and potential dependency.

The authors suggest future research should focus on optimizing the design for portability and comfort, and validating the system's effectiveness in more complex and real-world scenarios.

Summary updated

BibTeX
@inproceedings{xu2025adjustable,
  author = {Tianyu Xu and Zhenkun Sun and Sujiao Li and Hongyan Tang and Yanbin Zhang and Raymond Kai‐Yu Tong and Qiaoling Meng and Hongliu Yu},
  title = {{Adjustable-Stiffness Hip Exoskeleton with Flexible Energy-Storage Module for 3D Gait Correction}},
  booktitle = {Machines},
  year = {2025},
  doi = {10.3390/machines13100959},
}
Figure