Absolute quantitation of myocardial activity in phantoms

A.j. Da Silva, H.r. Tang, Ming C. Wu, B.h. HasegawaIEEE Transactions on Nuclear Science 1999 · DOI

Abstract

We have developed a new technique for compensating myocardial SPECT images for partial volume errors using co-registered X-ray CT images. The CT-derived myocardial mass defines a template that can be assigned a unit activity and mathematically projected with a realistic physical model of the radionuclide imaging process, including non-ideal collimation and incorporating an object-specific attenuation map from CT. The template projections then are reconstructed using a SPECT reconstruction algorithm to obtain a pixel-by-pixel partial-volume correction for the myocardial SPECT image. Experiments in phantoms demonstrate that this technique substantially improves the absolute quantitation of myocardial radionuclide concentration, reducing the accuracy error from approximately 50% to less than 8%. This method also can by used for correcting background effects such as "spill-in" of background counts from uptake in the liver.
BibTeX
@article{silva1999absolute,
  author = {A.J. Da Silva and H.R. Tang and Ming C. Wu and B.H. Hasegawa},
  title = {{Absolute quantitation of myocardial activity in phantoms}},
  journal = {IEEE Transactions on Nuclear Science},
  year = {1999},
  doi = {10.1109/23.775595},
}

Loading related content...

Figure