Absolute in vivo quantitation of myocardial activity

A.j. Da Silva, H.r. Tang, Kenneth H. Wong, Ming C. Wu, Michael W. Dae, B.h. HasegawaIEEE Transactions on Nuclear Science 2000 · DOI

Abstract

Quantitation of myocardial SPECT images corrected for attenuation underestimates the true radionuclide content due to partial volume errors. To measure radionuclide uptake more accurately, the authors have developed a technique to compensate these images for partial volume errors using coregistered X-ray CT images. The CT image is used to define a template that approximates the geometrical extent of the myocardium. Once defined, the template is assigned unit activity and is mathematically projected using a realistic physical model of the radionuclide imaging process. These projections are then reconstructed and used to compensate the SPECT image for partial volume errors. The method was tested in a porcine model of myocardial perfusion using Tc-99m sestamibi. With attenuation correction alone, the in vivo activity concentration in the porcine myocardium had an error in the range -40% to -60% compared to the true activity concentration. Correcting for partial volume errors, the in vivo activity concentration was determined to within 10% of the true value. By also correcting for partial volume errors, the in vivo activity concentration was determined to within 10% of the true value.