Absolute distance measurement in a combined-dispersive interferometer using a femtosecond pulse laser

Hanzhong Wu, Fumin Zhang, Fei Meng, Tingyang Liu, Jianshuang Li, Liang Pan, Xinghua QuMeasurement 2015 · DOI

Abstract

In this paper, a ranging system using dispersive interferometry is developed with a femtosecond pulse laser, aiming to eliminate the measurement dead zones by using a greatly unbalanced Mach–Zehnder interferometer. The distance can be measured by the frequency of the spectral modulation. We indicate that the integer number of the pulse-to-pulse length can be determined by changing the repetition frequency. In the short distance measurement, the results show an agreement within 1.5 μm compared with an incremental He-Ne laser in the 1 m measurement range. We do large-scale experiments on a long optical rail using a typical Michelson interferometer, and an agreement well within 25 μm is obtained in a range up to 75 m, corresponding to a relative precision of 3.3 × 10−7. Additionally, we experimentally optimize the system set-up to minimize the measurement uncertainty.
BibTeX
@inproceedings{wu2015absolute,
  author = {Hanzhong Wu and Fumin Zhang and Fei Meng and Tingyang Liu and Jianshuang Li and Liang Pan and Xinghua Qu},
  title = {{Absolute distance measurement in a combined-dispersive interferometer using a femtosecond pulse laser}},
  booktitle = {Measurement},
  year = {2015},
  doi = {10.1088/0957-0233/27/1/015202},
}

Loading related content...

Figure