NeurIPS 2026Coming to NeurIPS?Explore papers, orals & directions

A Simultaneous Measurement of Opposite-Phase-Encoding EPI in a Single fMRI session for the Reduction of Acquisition Time and SAR at 7T

Seong Dae Yun, Erhan Genc, N. Jon ShahProceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition 2024

Abstract

In high-resolution fMRI using EPI, geometric distortions typically seen in reconstructed images significantly hinder the accurate mapping of activated voxels. One method to correct for distortions is to acquire EPI data with an opposite phase-encoding direction. However, this method is usually implemented with an additional run of the same protocol, leading to a redundant measurement of parallel imaging calibration scans. Here, we present an EPI scheme that measures the opposite-direction data in a single fMRI session, substantially reducing the total acquisition time. We demonstrate more accurate functional mapping with the distortion correction in submillimetre whole-brain visual fMRI at 7T.
BibTeX
@inproceedings{yun2024simultaneous,
  author = {Seong Dae Yun and Erhan Genc and N. Jon Shah},
  title = {{A Simultaneous Measurement of Opposite-Phase-Encoding EPI in a Single fMRI session for the Reduction of Acquisition Time and SAR at 7T}},
  booktitle = {Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition},
  year = {2024},
  doi = {10.58530/2023/1161},
}

Loading related content...

Figure