A Sensing Strategy Combining Kinematic Model With FBG Sensors for Continuum Robots
Abstract
Continuum manipulators (CMs) are commonly used in surgical robots due to their compact size and flexibility. However, these features also lead to low rigidity and limited accuracy. To enhance localization accuracy, this paper proposes a particle filter-based localization algorithm that combines kinematic and fiber Bragg grating (FBG) sensors data. Firstly, a tiny (Φ 4 mm) CM embedded with FBG sensors was designed, its kinematic model was derived based on the constant curvature model, and its measurement results were obtained by calculating curvature, interpolation, and integration according to the FBG sensor data. Subsequently, an adaptive weighted particle filter algorithm is proposed, which can prevent erroneous outputs when the kinematic model fails A series of experiments were conducted to demonstrate the proposed method. Bending experiments showed that, when the model is valid, the proposed method reduces error by 50% compared to measurement alone, with an error consistently under 0.8 mm across experimental conditions regardless of model validity. In trajectory ablation experiments on phantoms, the proposed method aligned more accurately with the ablation path.
BibTeX
@article{hou2025sensing,
author = {Licheng Hou and Xingwei Zhao and Teru Chen and Yuhao Zhang and Qing Ling and Bo Tao},
title = {{A Sensing Strategy Combining Kinematic Model With FBG Sensors for Continuum Robots}},
journal = {IEEE Transactions on Instrumentation and Measurement},
year = {2025},
doi = {10.1109/tim.2025.3561417},
}Ask Althea to put this paper to the test.