A Rapid One-Pot Workflow for Sensitive Microscale Phosphoproteomics

Gul Muneer, Ciao‐Syuan Chen, T D Lee, Bo‐Yu Chen, Yu‐Ju ChenJournal of Proteome Research 2024 · DOI

TL;DR

Researchers have developed a streamlined, one-pot phosphoproteomics workflow (SOP-Phos) that significantly boosts sensitivity and reduces sample input requirements. This new method, combined with data-independent acquisition mass spectrometry (DIA-MS) and a tailored spectral library, enables the identification of thousands of phosphopeptides from microgram-level cell lysates and even down to 2,500 cells, paving the way for deeper insights into complex biological processes and disease mechanisms.

What Problem Does This Paper Address?

Phosphoproteomics, the study of protein phosphorylation, is crucial for understanding cellular signaling and disease. However, it has historically lagged behind general proteomics in terms of sensitivity, largely due to the low abundance of phosphopeptides and the complex, multi-step sample preparation required. These lengthy protocols lead to significant sample loss, demanding substantial starting material (hundreds of micrograms of protein) and limiting its application to challenging samples like rare cell populations or small biopsies. The authors aimed to overcome these limitations by developing a rapid, simplified, and highly sensitive workflow for microscale phosphoproteomics.

Methods

The core innovation is the SOP-Phos (Simple and Rapid One-Pot Phosphoproteomics) workflow. Key features include:

  • DDM Coating: Tubes are pre-coated with n-dodecyl β-D-maltoside (DDM) to minimize adsorptive losses of precious peptides and proteins. This is a crucial step for low-input samples.
  • One-Pot Lysis, Reduction, and Alkylation: Sodium deoxycholate (SDC) is used as a lysis buffer, allowing for concurrent cell lysis, protein reduction, and alkylation in a single step, eliminating the need for protein precipitation and detergent removal.
  • Rapid Digestion: Trypsin digestion time is significantly shortened from overnight to 1-2 hours by increasing the enzyme-to-substrate ratio.
  • Direct Phosphopeptide Enrichment: TiO₂ beads are used for phosphopeptide enrichment directly from the digested sample, bypassing intermediate cleanup steps.
  • Data-Independent Acquisition (DIA-MS): This mass spectrometry acquisition method was employed for its high sensitivity, specificity, and reproducibility.
  • Sample-Size Comparable Spectral Library: A custom spectral library, built from data acquired with similar sample input levels, was used to enhance the identification of phosphopeptides, particularly those present at low abundance.

Datasets

The study utilized PC9, CL68, H1975, and H3255 lung cancer cell lines. Different sample inputs were tested, ranging from 5 µg down to 0.5 µg of cell lysate, and from 50,000 down to 2,500 cells. Spectral libraries were constructed from both DDA (Data-Dependent Acquisition) and DIA data.

Key Insights and Ideas

  • Minimizing Sample Loss is Paramount: The study strongly emphasizes that reducing sample loss through streamlined protocols and surface passivation (DDM coating) is critical for achieving high sensitivity in low-input phosphoproteomics.
  • One-Pot Simplicity Drives Efficiency: Combining multiple steps into a single tube dramatically reduces processing time and potential for sample loss, making the workflow more accessible.
  • DIA-MS and Targeted Libraries are a Powerful Combination: While DIA-MS offers inherent advantages, pairing it with a spectral library that matches the sample input level significantly boosts the identification of low-abundance phosphopeptides.
  • Rapid Digestion is Feasible and Effective: Shortening digestion times from overnight to a few hours does not compromise phosphopeptide identification and greatly accelerates the overall workflow.
  • EGFR-TKI Resistance is a Complex Interplay: The method successfully revealed intricate signaling network alterations, particularly involving the Hippo, EGFR, and ERBB pathways, in EGFR-TKI resistant lung cancer cells, offering new avenues for therapeutic intervention.

Key Results

  • High Phosphopeptide Coverage: The SOP-Phos-DIA workflow identified 33,787 ± 670 to 22,070 ± 861 phosphopeptides from 5 µg to 0.5 µg of cell lysate.
  • Low-Cell Input Success: The method identified 30,433 ± 284 to 6,548 ± 21 phosphopeptides from as few as 50,000 to 2,500 cells.
  • Improved Reproducibility: The workflow demonstrated high reproducibility with low missing values (<1%) and coefficients of variation (8-10% CV).
  • Time Efficiency: The entire workflow can be completed in 3-4 hours.
  • Biological Insights: The study successfully mapped key lung cancer signaling sites, including EGFR autophosphorylation sites (Y1197/Y1172), and provided mechanistic insights into EGFR-TKI resistance by revealing alterations in multipathway signaling cascades.

Limitations and Future Work

  • DDM Interference: While DDM coating was effective, the authors noted that DDM itself was detected in the mass spectrometry data, though at trace levels. They also excluded DDM as a lysis buffer due to potential LC-column damage at higher concentrations.
  • Spectral Library Optimization: The study highlights the importance of sample-size comparable spectral libraries. Future work could focus on further optimizing library construction strategies for even lower input amounts or different cell types.
  • Broader Applicability: While demonstrated on lung cancer cells, further validation across diverse cell types and biological contexts would be beneficial.
  • Integration with Miniaturized Devices: The authors suggest that this workflow could be integrated into miniaturized devices, paving the way for single-cell phosphoproteomics.

Summary updated

Abstract

Compared to advancements in single-cell proteomics, phosphoproteomics sensitivity has lagged behind due to low abundance, complex sample preparation, and substantial sample input requirements. We present a simple and rapid one-pot phosphoproteomics workflow (SOP-Phos) integrated with data-independent acquisition mass spectrometry (DIA-MS) for microscale phosphoproteomic analysis. SOP-Phos adapts sodium deoxycholate based one-step lysis, reduction/alkylation, direct trypsinization, and phosphopeptide enrichment by TiO<sub>2</sub> beads in a single-tube format. By reducing surface adsorptive losses via utilizing <i>n</i>-dodecyl β-d-maltoside precoated tubes and shortening the digestion time, SOP-Phos is completed within 3-4 h with a 1.4-fold higher identification coverage. SOP-Phos coupled with DIA demonstrated >90% specificity, enhanced sensitivity, lower missing values (<1%), and improved reproducibility (8%-10% CV). With a sample size-comparable spectral library, SOP-Phos-DIA identified 33,787 ± 670 to 22,070 ± 861 phosphopeptides from 5 to 0.5 μg cell lysate and 30,433 ± 284 to 6,548 ± 21 phosphopeptides from 50,000 to 2,500 cells. Such sensitivity enabled mapping key lung cancer signaling sites, such as EGFR autophosphorylation sites Y1197/Y1172 and drug targets. The feasibility of SOP-Phos-DIA was demonstrated on EGFR-TKI sensitive and resistant cells, revealing the interplay of multipathway Hippo-EGFR-ERBB signaling cascades underlying the mechanistic insight into EGFR-TKI resistance. Overall, SOP-Phos-DIA is an efficient and robust protocol that can be easily adapted in the community for microscale phosphoproteomic analysis.
BibTeX
@article{muneer2024rapid,
  author = {Gul Muneer and Ciao‐Syuan Chen and T D Lee and Bo‐Yu Chen and Yu‐Ju Chen},
  title = {{A Rapid One-Pot Workflow for Sensitive Microscale Phosphoproteomics}},
  journal = {Journal of Proteome Research},
  year = {2024},
  doi = {10.1021/acs.jproteome.3c00862},
}