A Quadruple-Sweep Bioimpedance Sensing Method for Arterial Stenosis Detection

Enayetur Rahman, Panos Liatsis, Zaheer Q. Hashim, Panayiotis A. Kyriacou, Iasonas F. TriantisIEEE Access 2024 · OpenReview

Abstract

Current carotid atherosclerosis diagnostic protocols do not feature techniques that would allow for early or frequent medical examinations, leaving a significant number of asymptomatic carotid stenosis cases undetected and often leading to strokes. The key challenge is that current diagnostics are highly operator-dependent. In this work we used idealised biological models to demonstrate a new rapid, potentially inexpensive and operator-independent diagnostic method, aimed at detecting whether a stenosis exists, rather than seeking to be accurately quantifying or localising it. An array of electrodes was used to obtain sequential bioimpedance values over the skin, through a novel scanning technique, covering an area over the artery of interest. FEM simulations, verified through in-vitro experiments on gelatine phantoms, were used to validate the method. The final results, obtained through image processing algorithms, were in the form of planar bio-impedance maps and were successful both in identifying arterial features and detecting the presence of stenoses of different sizes, and the overall accuracy of detecting stenoses in a vessel without bifurcation is 97.78% and with bifurcation is 91.11%. The results could also be used to indicate the artery’s relative orientation to the sensor, eliminating the need for manual alignment by a specialist operator. Therefore, this method shows promise for routine medical examination, either in primary care, or even at home, to indicate whether a patient would require further, more detailed examinations at a specialist clinic.

What Problem Does This Paper Address?

Carotid artery disease is a leading cause of stroke, yet it is often asymptomatic until a major medical event occurs. The current "gold standard" for screening is ultrasound imaging. While effective, ultrasound is highly operator-dependent—the quality of the scan depends entirely on the skill of the person holding the probe. This leads to high costs and error rates as high as 18%, meaning many patients aren't screened until they already show symptoms.

Comparison between traditional operator-dependent ultrasound (a) and the proposed operator-independent bioimpedance patch (b).

The authors aim to bridge this gap by creating a diagnostic tool that is inexpensive, portable, and, most importantly, operator-independent. They propose a device that can be placed on the neck and automatically detect if a blockage exists, regardless of the exact alignment.

Methods: Localised Planar Bioimpedance Mapping

The team introduced a technique called Localised Planar Bioimpedance Mapping (PBiM). This approach uses an array of electrodes placed on the skin to obtain sequential bioimpedance values.

  1. The Hardware: They used an array of electrodes (20x30 for simulations, 8x16 for physical experiments) placed on the skin.
  2. The "Quadruple-Sweep": Instead of a single measurement, the system performs four sequential scans: vertical, horizontal, diagonal-left, and diagonal-right. This uses a "shifting tetrapole" configuration, where four neighboring electrodes work together (two to inject a tiny current, two to measure the resulting voltage).
  3. Validation: The method was tested using Finite Element Method (FEM) simulations in COMSOL and verified with in-vitro gelatine phantoms. These phantoms mimic the electrical conductivity of human skin, fat, and blood.
Experimental setup showing the electrode array over (b) a straight vessel and (c) a bifurcated vessel with stenosis.
  1. Image Processing: The raw impedance data is converted into a 2D map. The researchers used Otsu’s thresholding to separate the vessel from surrounding tissue and a k-means++ clustering algorithm to specifically identify the presence of a stenosis (blockage) within the vessel.

Key Insights and Ideas

  • Orientation Independence: By combining four different scanning directions, the system can calculate the artery's orientation relative to the sensor. This removes the need for a specialist to manually align a probe with the artery.
  • Feature over Quantity: Unlike Electrical Impedance Tomography (EIT), which tries to reconstruct a high-resolution 3D image, this method focuses on feature detection. It asks, "Is there a vessel here, and is it blocked?" rather than trying to create a perfect picture.
  • Bifurcation Matters: Most blockages occur where the carotid artery splits (the bifurcation). The study specifically modeled these Y-junctions, proving the method works in the most clinically relevant areas of the neck.

Key Results

The system demonstrated high efficacy in detecting arterial features and stenoses:

  • High Accuracy: The system achieved a 97.78% accuracy in detecting stenosis in straight vessels and 91.11% accuracy in bifurcated vessels.
  • Precision Orientation: The algorithm could estimate the vessel's angle within ±3° of its actual position.
  • Low-Grade Detection: Even when blockages were relatively small (20%), the system showed a strong ability to distinguish them from healthy vessels.

Limitations and Future Work

While promising, the study used idealized biological models. The current phantoms and simulations do not account for complex anatomy, such as muscle layers, multiple neighboring veins, and varying skin thicknesses. Additionally, patient breathing or minor movements could introduce noise into the electrical readings in a real-world setting. Ensuring consistent electrical contact between the array and the curved surface of a human neck remains a mechanical challenge.

Future work will involve more complex "anatomically correct" models and moving toward clinical trials with human subjects.

Summary updated

Abstract

Current carotid atherosclerosis diagnostic protocols do not feature techniques that would allow for early or frequent medical examinations, leaving a significant number of asymptomatic carotid stenosis cases undetected and often leading to strokes. The key challenge is that current diagnostics are highly operator-dependent. In this work we used idealised biological models to demonstrate a new rapid, potentially inexpensive and operator-independent diagnostic method, aimed at detecting whether a stenosis exists, rather than seeking to be accurately quantifying or localising it. An array of electrodes was used to obtain sequential bioimpedance values over the skin, through a novel scanning technique, covering an area over the artery of interest. FEM simulations, verified through in-vitro experiments on gelatine phantoms, were used to validate the method. The final results, obtained through image processing algorithms, were in the form of planar bio-impedance maps and were successful both in identifying arterial features and detecting the presence of stenoses of different sizes, and the overall accuracy of detecting stenoses in a vessel without bifurcation is 97.78% and with bifurcation is 91.11%. The results could also be used to indicate the artery’s relative orientation to the sensor, eliminating the need for manual alignment by a specialist operator. Therefore, this method shows promise for routine medical examination, either in primary care, or even at home, to indicate whether a patient would require further, more detailed examinations at a specialist clinic.