A Power and Data Link for a Wireless-Implanted Neural Recording System

Alexander M. Rush, Philip R. TroykIEEE Trans. Biomedical Engineering 2012 · DOI

Abstract

A wireless cortical neural recording system with a miniature-implanted package is needed in a variety of neuroscience and biomedical applications. Toward that end, we have developed a transcutaneous two-way communication and power system for wireless neural recording. Wireless powering and forward data transmission (into the body) at 1.25 Mbps is achieved using a frequency-shift keying modulated class E converter. The reverse telemetry (out of the body) carrier frequency is generated using an integer-N phase-locked loop, providing the necessary wideband data link to support simultaneous reverse telemetry from multiple implanted devices on separate channels. Each channel is designed to support reverse telemetry with a data rate in excess of 3 Mbps, which is sufficient for our goal of streaming 16 channels of raw neural data. We plan to incorporate this implantable power and telemetry system in a 1-cm diameter single-site cortical neural recording implant.
BibTeX
@article{rush2012power,
  author = {Alexander M. Rush and Philip R. Troyk},
  title = {{A Power and Data Link for a Wireless-Implanted Neural Recording System}},
  journal = {IEEE Transactions on Biomedical Engineering},
  year = {2012},
  doi = {10.1109/tbme.2012.2214385},
}

Loading related content...

Figure